Influence of Magnetic Field Strength on Intravoxel Incoherent Motion Parameters in Diffusion MRI of the Calf
Tomography · 2024
DOI: 10.3390/tomography10050059 ↗Kurz gefasst
Fachartikel in der Zeitschrift Tomography, 2024. Verfasst von Tamara Alice Bäuchle und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany. Hauptthemen laut MEDLINE: Diffusion Magnetic Resonance Imaging, Muscle, Skeletal. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Background
The purpose of this study was to investigate the dependence of Intravoxel Incoherent Motion (IVIM) parameters measured in the human calf on B0.
Methods
Diffusion-weighted image data of eight healthy volunteers were acquired using five b-values (0–600 s/mm2) at rest and after muscle activation at 0.55 and 7 T. The musculus gastrocnemius mediale (GM, activated) was assessed. The perfusion fraction f and diffusion coefficient D were determined using segmented fits. The dependence on field strength was assessed using Student’s t-test for paired samples and the Wilcoxon signed-rank test. A biophysical model built on the three non-exchanging compartments of muscle, venous blood, and arterial blood was used to interpret the data using literature relaxation times.
Results
The measured perfusion fraction of the GM was significantly lower at 7 T, both for the baseline measurement and after muscle activation. For 0.55 and 7 T, the mean f values were 7.59% and 3.63% at rest, and 14.03% and 6.92% after activation, respectively. The biophysical model estimations for the mean proton-density-weighted perfusion fraction were 3.37% and 6.50% for the non-activated and activated states, respectively.
Conclusions
B0 may have a significant effect on the measured IVIM parameters. The blood relaxation times suggest that 7 T IVIM may be arterial-weighted whereas 0.55 T IVIM may exhibit an approximately equal weighting of arterial and venous blood.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany.
ORCID-Profile: Christoph Martin Stuprich, Frederik Bernd Laun
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Bäuchle, T. A., Stuprich, C. M., Loh, M., Nagel, A. M., Uder, M., & Laun, F. B. (2024). Influence of Magnetic Field Strength on Intravoxel Incoherent Motion Parameters in Diffusion MRI of the Calf. Tomography. https://doi.org/10.3390/tomography10050059
Vancouver
Bäuchle TA, Stuprich CM, Loh M, Nagel AM, Uder M, Laun FB. Influence of Magnetic Field Strength on Intravoxel Incoherent Motion Parameters in Diffusion MRI of the Calf. Tomography. 2024. doi:10.3390/tomography10050059
BibTeX
@article{Buchle2024,
author = {Bäuchle, Tamara Alice and Stuprich, Christoph Martin and Loh, Martin and Nagel, Armin Michael and Uder, Michael and Laun, Frederik Bernd},
title = {Influence of Magnetic Field Strength on Intravoxel Incoherent Motion Parameters in Diffusion MRI of the Calf},
journal = {Tomography},
year = {2024},
doi = {10.3390/tomography10050059}
}