EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Development and Testing of NDIR-Based Rapid Greenhouse Gas Detection Device for Dairy Farms

Qianwen Li, Yongkang He, Kaixuan Zhao, Jiangtao Ji, Hongzhen Li, Jeffrey M. Bewley

Sustainability · 04.03.2024 · 16 (5) · 2131

DOI: 10.3390/su16052131 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sustainability, online erschienen am 04.03.2024. Verfasst von Qianwen Li und 5 weiteren Autor:innen. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the National Key R&D Plan Key projects, International Science and Technology Cooperation Project of Henan Province Key Research and Development Projects, University science and technology innovation talent project of Henan Province.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

As greenhouse gas emissions from dairy farms are on the rise, effective monitoring of these emissions has emerged as a crucial tool for assessing their environmental impacts and promoting sustainable development. Most of the existing studies on GHGs from dairy farms involve stationary detections with long response times and high costs. In this study, a greenhouse gas detection system was constructed based on NDIR technology using a single broadband light source and a four-channel thermopile detector for the detection of CH4, N2O, and CO2; the detection range of CH4 was 0~100 ppm; that of N2O was 0~500 ppm; and that of CO2 was 0~20%. After the concentration calibration, the cross-interference between the gas measurement channels was studied, and the least-squares method was used to correct the interference between the three gases. The experimental results showed that the full-range deviation of the detection device was lower than 0.81%, the repeatability was lower than 0.39%, the stability was lower than 0.61%, and the response time was lower than 10 s. This study also carried out on-site testing in Luoyang Shengsheng Ranch (Luoyang, China), and the results show that the error between this device and the PTM600 portable gas analyzer is within 9.78%, and the dynamic response time of this device is within 16 s, at which point the content of greenhouse gases in dairy farms can be measured quickly and accurately. The objective of this study is to enhance the precision and effectiveness of greenhouse gas (GHG) emissions monitoring from dairy farms, thereby contributing to environmental protection and sustainable development goals. By achieving this, we aim to facilitate societal progress towards a greener and low-carbon future.

Förderung

  • the National Key R&D Plan Key projects
  • International Science and Technology Cooperation Project of Henan Province Key Research and Development Projects
  • University science and technology innovation talent project of Henan Province

Zitieren

APA

Li, Q., He, Y., Zhao, K., Ji, J., Li, H., & Bewley, J. M. (2024). Development and Testing of NDIR-Based Rapid Greenhouse Gas Detection Device for Dairy Farms. Sustainability, 16(5), 2131. https://doi.org/10.3390/su16052131

Vancouver

Li Q, He Y, Zhao K, Ji J, Li H, Bewley JM. Development and Testing of NDIR-Based Rapid Greenhouse Gas Detection Device for Dairy Farms. Sustainability. 2024;16(5):2131. doi:10.3390/su16052131

BibTeX
@article{Li2024,
  author = {Li, Qianwen and He, Yongkang and Zhao, Kaixuan and Ji, Jiangtao and Li, Hongzhen and Bewley, Jeffrey M.},
  title = {Development and Testing of NDIR-Based Rapid Greenhouse Gas Detection Device for Dairy Farms},
  journal = {Sustainability},
  year = {2024},
  volume = {16},
  number = {5},
  pages = {2131},
  doi = {10.3390/su16052131}
}