Impact of Controlled Superheated Steam Conditions on a Commercial Peel-Pouch Sterile Barrier Systems: A Pilot Study

William Leiva, Atila Nozari

Standards · 06.08.2026 · 6 (3) · 30

DOI: 10.3390/standards6030030 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Standards, online erschienen am 06.08.2026. Verfasst von William Leiva und 1 weiteren Autor:innen. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: AAMI Foundation Kilmer Grant, 3M Healthcare.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Saturated steam remains the primary sterilization modality in healthcare facilities globally; however, superheated steam may compromise sterile barrier integrity. Limited data characterize how superheated steam affects peel-pouch sterile barriers under controlled conditions. This pilot study evaluated peel-pouch performance under saturated steam control and experimental superheated steam conditions. Seventy-two samples were exposed to four exposure scenarios using an ISO 18472:2018 compliant BIER vessel: control and experimental treatment at both 121 °C and 132 °C (N = 18 per group). Outcomes included visual inspection (cracking, discoloration, peel-off), bubble emission testing and peel strength testing. Fisher exact tests with Bonferroni correction evaluated categorical outcomes; two-way ANOVA evaluated peel force data. Visual damage showed no statistically significant difference across conditions (Fisher–Freeman–Halton exact test, p = 0.422), ranging from 5.6% in controls to 22.2% at 121 °C experimental conditions (OR = 4.85, p = 0.038). Pairwise odds ratio estimates suggested a possible increase in visible damage under experimental conditions; however, the overall comparison was not statistically significant. Bubble emission testing revealed highly significant differences (p = 0.0014), with experimental conditions showing 71.43% failure at 121 °C versus 11.11% in controls. Peel strength was dramatically reduced in experimental samples (47.4% reduction; F = 154.83, p < 0.0001), independent of temperature. Visual inspection failed to detect barrier compromises evident through functional and mechanical testing. In controlled laboratory conditions using a single peel-pouch brand, exposure to controlled superheated steam conditions was associated with significant barrier damage. These findings underscore the need for comprehensive barrier testing beyond visual inspection in sterilization validation.

Förderung

  • AAMI Foundation Kilmer Grant
  • 3M Healthcare

Zitieren

APA

Leiva, W., & Nozari, A. (2026). Impact of Controlled Superheated Steam Conditions on a Commercial Peel-Pouch Sterile Barrier Systems: A Pilot Study. Standards, 6(3), 30. https://doi.org/10.3390/standards6030030

Vancouver

Leiva W, Nozari A. Impact of Controlled Superheated Steam Conditions on a Commercial Peel-Pouch Sterile Barrier Systems: A Pilot Study. Standards. 2026;6(3):30. doi:10.3390/standards6030030

BibTeX
@article{Leiva2026,
  author = {Leiva, William and Nozari, Atila},
  title = {Impact of Controlled Superheated Steam Conditions on a Commercial Peel-Pouch Sterile Barrier Systems: A Pilot Study},
  journal = {Standards},
  year = {2026},
  volume = {6},
  number = {3},
  pages = {30},
  doi = {10.3390/standards6030030}
}
Kostenlos starten →