Effects of Illumination and Resonance Spectral Conditions on the Sensing Performance of Guided-Mode Resonance Sensors

Shih-Chuan Huang, Cheng-Sheng Huang

Sensors · 19.11.2026 · 26 (22) · 7332

DOI: 10.3390/s26227332 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 19.11.2026. Verfasst von Shih-Chuan Huang und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Education, National Science and Technology.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Intensity-based interrogation is a simple, inexpensive readout strategy for guided-mode resonance (GMR) sensors; however, its performance depends on the overlap between illumination and resonance spectra. We developed an overlap integral model that quantifies a GMR sensor’s transmission intensity as a function of illumination wavelength, illumination and resonance full widths at half maximum (FWHMs), resonance efficiency, and wavelength offset from the resonance dip. By using normalized intensity sensitivity as a figure of merit, we demonstrated that positioning the illumination wavelength on the linear flank of the resonance yields higher sensitivity than centering it on the resonance dip for small GMR resonance FWHMs, especially under small refractive-index changes. Sensitivity increases with resonance efficiency and decreases with illumination FWHM but varies non-monotonically with resonance FWHM, revealing trade-offs among sensitivity, signal magnitude, and dynamic range. These predictions were experimentally validated using a GMR-integrated polydimethylsiloxane microfluidic chip interrogated by a tunable-FWHM monochromator and 0–10% sucrose solutions. The measured transmittance changes reproduced the predicted behaviors and the sensitivity gain from reducing the illumination linewidth from 3 to 1 nm. Overall, the results guide the selection of illumination offset, source linewidth, and resonance linewidth in the design of compact, spectrometer-free GMR sensors for point-of-care diagnostics.

Förderung

  • Ministry of Education
  • National Science and Technology

Zitieren

APA

Huang, S., & Huang, C. (2026). Effects of Illumination and Resonance Spectral Conditions on the Sensing Performance of Guided-Mode Resonance Sensors. Sensors, 26(22), 7332. https://doi.org/10.3390/s26227332

Vancouver

Huang S, Huang C. Effects of Illumination and Resonance Spectral Conditions on the Sensing Performance of Guided-Mode Resonance Sensors. Sensors. 2026;26(22):7332. doi:10.3390/s26227332

BibTeX
@article{Huang2026,
  author = {Huang, Shih-Chuan and Huang, Cheng-Sheng},
  title = {Effects of Illumination and Resonance Spectral Conditions on the Sensing Performance of Guided-Mode Resonance Sensors},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {22},
  pages = {7332},
  doi = {10.3390/s26227332}
}
Kostenlos starten →