Reference-Assisted FBG Sensing for Simultaneous Carotid–Radial Pulse-Wave Measurement and PWV Estimation
Sensors · 30.10.2026 · 26 (21) · 6879
DOI: 10.3390/s26216879 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 30.10.2026. Verfasst von Chang Hyun Park und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cogno-Mechatronics Engineering, Pusan National University, Busan 46241, Republic of Korea. Hauptthemen laut MEDLINE: Pulse Wave Analysis, Radial Artery, Carotid Arteries. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Korea Health Industry Development Institute, Commercialization Promotion Agency for R&D Outcomes.
Abstract (Originalsprache)
Simultaneous pulse-wave sensing at the carotid and radial arteries is challenging because the two sites require different mechanical coupling conditions and manually held carotid probes are susceptible to contact pressure and motion disturbances. This study presents a reference-assisted fiber Bragg grating (FBG) system comprising a wristband-type radial unit and a stamp-type carotid unit containing mechanically decoupled measurement and reference FBGs. Reference FBG regression (RFR) was used to suppress components shared by the carotid channels. In a 41.8 s recording from one male participant in his twenties, the reference and carotid measurement channels were essentially uncorrelated within the pulse-wave band (r = −0.05). After signal mean removal, RFR reduced power below 1 Hz by 74.3% without attenuating the pulse-band power (+11.4%), increasing the relative-power signal quality index from 0.174 to 0.458. Pulse transit time (PTT) was estimated by foot-centered windowed correlation. Analysis of 54 cardiac cycles yielded a representative PTT of 0.055 s. With an effective path length of 0.48 m, the corresponding pulse-wave velocity (PWV) was 8.7 m/s, with a within-record bootstrap 95% interval of 7.7–9.9 m/s. Estimates from eight non-overlapping 5 s segments were 9.1 ± 1.8 m/s. RFR produced lower segment-to-segment PTT variability than the baseline processing methods (coefficient of variation, 20.3% versus 34.6–46.7%). These results demonstrate the technical feasibility of reference-assisted FBG sensing for simultaneous carotid–radial pulse-wave acquisition and preliminary PWV estimation. Further validation requires higher-rate acquisition, repeated measurements, a larger cohort, and comparison with a reference PWV device.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cogno-Mechatronics Engineering, Pusan National University, Busan 46241, Republic of Korea.
- School of Transdisciplinary Engineering, Pusan National University, Busan 46241, Republic of Korea.
- ARK Inc., Busan 46241, Republic of Korea.
- Engineering Research Center for Color-Modulated Extra-Sensory Perception Technology, Pusan National University, Busan 46241, Republic of Korea.
- The Crystal Bank Research Institute, Pusan National University, Busan 46241, Republic of Korea.
ORCID-Profile: Chang Hyun Park, Minju Jeong, Jeesu Kim, Chang-Seok Kim, Hwidon Lee
Förderung
- National Research Foundation of Korea
- Korea Health Industry Development Institute
- Commercialization Promotion Agency for R&D Outcomes
Zitieren
APA
Park, C. H., Jeong, M., Jeong, S., Hur, M., Lee, Y., Kim, H. H., Kim, J., Kim, C. S., & Lee, H. (2026). Reference-Assisted FBG Sensing for Simultaneous Carotid–Radial Pulse-Wave Measurement and PWV Estimation. Sensors, 26(21), 6879. https://doi.org/10.3390/s26216879
Vancouver
Park CH, Jeong M, Jeong S, Hur M, Lee Y, Kim HH, et al. Reference-Assisted FBG Sensing for Simultaneous Carotid–Radial Pulse-Wave Measurement and PWV Estimation. Sensors. 2026;26(21):6879. doi:10.3390/s26216879
BibTeX
@article{Park2026,
author = {Park, Chang Hyun and Jeong, Minju and Jeong, Seongmun and Hur, Min and Lee, Yoonji and Kim, Hyung-Hoi and Kim, Jeesu and Kim, Chang-Seok and Lee, Hwidon},
title = {Reference-Assisted FBG Sensing for Simultaneous Carotid–Radial Pulse-Wave Measurement and PWV Estimation},
journal = {Sensors},
year = {2026},
volume = {26},
number = {21},
pages = {6879},
doi = {10.3390/s26216879}
}
EUVIMED