EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study

Enver Salkim iD, Andreas Demosthenous iD

Sensors · 30.10.2026 · 26 (21) · 6870

DOI: 10.3390/s26216870 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 30.10.2026. Verfasst von Enver Salkim und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronic and Automation, Muş Alparslan University, 49100 Muş, Türkiye. Hauptthemen laut MEDLINE: Electric Impedance, Cochlear Implants, Computer Simulation, Cochlear Implantation. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Intraoperative tissue trauma during cochlear implant (CI) insertion remains a primary cause of residual hearing loss. Because insertion currently relies largely on clinician experience without direct visualization within the optically obscured inner ear, real-time guidance is critical for minimizing mechanical contact with delicate anatomical structures. This study evaluates multi-channel electrical impedance for real-time trajectory tracking and boundary proximity detection during automated electrode insertion. A computational finite-element model of the human cochlea was used to analyze complex impedance profiles across eight bipolar configurations (E1–E2 to E15–E16) and 25 insertion models (M1–M25), yielding 2592 evaluations across central, medial, and lateral trajectories under varying bulk tissue conductivities. At 1kHz, reactive contributions were negligible (X2400 Ω, >99.5% of total impedance). Bulk conductivity scaled baseline impedance magnitudes without substantially altering the spatial profile geometry. Across the three insertion trajectories, impedance exhibited a characteristic U-shaped evolution, initially decreasing as the electrode array (EA) advanced toward the scala tympani (ST) center before increasing near the terminal insertion region. Boundary proximity further increased impedance variation, with mean impedance values increasing by 24–38% for the medial trajectory and 25–35% for the lateral trajectory relative to the middle baseline. Localized impedance increases were observed for the medial pair E3–E4 (65% at M20) and the lateral pair E1–E2 (88% at M23). The differential impedance profiles indicate that boundary-induced proximity effects substantially exceed baseline spatial variations, demonstrating the sensitivity of multi-channel impedance measurements to electrode–wall proximity. These findings support the potential use of impedance signatures as a real-time feedback signal for boundary proximity detection, clinician guidance, and closed-loop trajectory adjustment during automated CI insertion. Recognizing the characteristic U-shaped impedance profile and its sensitivity to boundary proximity may provide valuable intraoperative feedback for monitoring electrode-array position and reducing the risk of boundary contact and associated mechanical trauma.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Electric ImpedanceCochlear ImplantsComputer SimulationCochlear ImplantationHumansFinite Element AnalysisCochleaElectrodes, Implanted

Stichwörter der Autor:innen

bio-impedance analysisboundary proximity sensingcochlear implant (CI)finite element modelingreal-time trajectory trackingsensor-guided surgery

Autor:innen und Institutionen

  • Department of Electronic and Automation, Muş Alparslan University, 49100 Muş, Türkiye.
  • Department of Electronic and Electrical Engineering, University College London, London WC1E 7JE, UK.

ORCID-Profile: Enver Salkim, Andreas Demosthenous

Zitieren

APA

Salkim, E., & Demosthenous, A. (2026). Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study. Sensors, 26(21), 6870. https://doi.org/10.3390/s26216870

Vancouver

Salkim E, Demosthenous A. Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study. Sensors. 2026;26(21):6870. doi:10.3390/s26216870

BibTeX
@article{Salkim2026,
  author = {Salkim, Enver and Demosthenous, Andreas},
  title = {Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {21},
  pages = {6870},
  doi = {10.3390/s26216870}
}
Kostenlos starten →