EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study

Titus Jayarathna iD, Caitlin Polley, Paul P Breen iD, Gaetano D Gargiulo iD, Anders Aneman iD

Sensors · 2026

DOI: 10.3390/s26196052 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Sensors, 2026. Verfasst von Titus Jayarathna und 4 weiteren Autor:innen. Beteiligte Einrichtung: The MARCS Institute for Brain, Behaviour and Development, Western Sydney University, Locked Bag 1797, Penrith, NSW 2751, Australia. Hauptthemen laut MEDLINE: Heart Rate, Wearable Electronic Devices, Respiratory Rate. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This pilot study evaluated the real-world accuracy of a novel wearable morphic sensor system for unobtrusive, continuous heart rate (HR) and respiratory rate (RR) monitoring in critically ill patients within an intensive care unit (ICU).

Methods

A hybrid piezoresistive/piezoelectric disk-shaped sensor was placed on patients’ backs to capture cardiorespiratory activity via forcecardiographic principles. Data were acquired from 15 ICU patients using a custom wireless recorder. The piezoelectric channel was processed using discrete wavelet transform and an adaptive peak detection algorithm. Derived HR and RR were compared against reference measurements from clinical patient monitors and a ventilator.

Results

Bland–Altman analysis showed a mean bias of −0.39 ± 1.99 beats per minute for HR and 0.06 ± 0.90 breaths per minute for RR. Linear regression revealed strong correlation with clinical monitors (HR R-squared = 0.982; RR R-squared = 0.966). The device’s RR performance matched the agreement between two independent clinical reference devices.

Conclusion

The proposed system met the pre-specified bias and standard deviation criteria for HR and RR monitoring in bed-bound ICU patients. Its skin-adhesive-free design enables unobtrusive, continuous vital sign surveillance without direct skin contact. These findings support ICU monitoring. Step-down ward and ambulatory applications remain to be validated in future studies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart RateWearable Electronic DevicesRespiratory RateHumansIntensive Care UnitsPilot ProjectsMonitoring, PhysiologicFemaleMaleMiddle AgedAlgorithmsAgedSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

ICU monitoringdiscrete wavelet transformforcecardiographic signalmorphic sensorsvital signswearable electronics

Autor:innen und Institutionen

  • The MARCS Institute for Brain, Behaviour and Development, Western Sydney University, Locked Bag 1797, Penrith, NSW 2751, Australia.
  • School of Engineering, Western Sydney University, Locked Bag 1797, Penrith, NSW 2751, Australia.
  • Department of Intensive Care, Liverpool Hospital, Elizabeth Street, Liverpool, NSW 2170, Australia.

ORCID-Profile: Titus Jayarathna, Paul P Breen, Gaetano D Gargiulo, Anders Aneman

Zitieren

APA

Jayarathna, T., Polley, C., Breen, P. P., Gargiulo, G. D., & Aneman, A. (2026). A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study. Sensors. https://doi.org/10.3390/s26196052

Vancouver

Jayarathna T, Polley C, Breen PP, Gargiulo GD, Aneman A. A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study. Sensors. 2026. doi:10.3390/s26196052

BibTeX
@article{Jayarathna2026,
  author = {Jayarathna, Titus and Polley, Caitlin and Breen, Paul P and Gargiulo, Gaetano D and Aneman, Anders},
  title = {A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study},
  journal = {Sensors},
  year = {2026},
  doi = {10.3390/s26196052}
}
Kostenlos starten →