Long-Term Validation of a Consumer- vs. Industrial-Grade Indoor Radon and Air Quality Monitor at a School in Utah
Sensors · 18.08.2026 · 26 (16) · 5230
DOI: 10.3390/s26165230 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 18.08.2026. Verfasst von John D Beard und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Public Health, Brigham Young University, Provo, UT 84602, USA. Die Arbeit nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIOSH CDC HHS, National Institute for Occupational Safety and Health.
Abstract (Originalsprache)
In recent years, the demand for indoor radon and air quality monitors has increased while more affordable consumer-grade monitors have become available. The aim of our study was to assess the validity of the less expensive, non-calibrated, consumer-grade Lüft (SunRADON LLC, Melbourne, FL, USA) compared to the more expensive, annually and manufacturer calibrated, industrial-grade Model 1028 XP (SunRADON LLC, Melbourne, FL, USA). We collected data about 24 h moving averages of barometric pressure, carbon dioxide (CO2), radon, relative humidity, temperature, and volatile organic compounds (VOC) for approximately 29 weeks in 2024 in each of six rooms in a school in Utah. The mean difference in the co-located Lüfts’ and Model 1028 XPs’ 24 h moving average radon was −0.05 pCi/L of air (95% confidence interval: −0.24, 0.14, p = 0.50) and the concordance correlation coefficient (CCC) was 0.92. The Lüfts’ and Model 1028 XPs’ 24 h moving averages were similar for barometric pressure (CCC: 0.87), but the Lüfts’ 24 h moving averages were lower than the Model 1028 XPs’ for CO2, relative humidity, and VOC and higher for temperature. The Lüft provided valid measurements of 24 h moving averages for barometric pressure and radon, but not for CO2, relative humidity, temperature, and VOC.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Public Health, Brigham Young University, Provo, UT 84602, USA.
- Department of Electrical and Computer Engineering, Brigham Young University, Provo, UT 84602, USA.
ORCID-Profile: John D Beard, Christopher Kitras, Philip B Lundrigan
Förderung
- NIOSH CDC HHS
- National Institute for Occupational Safety and Health
Zitieren
APA
Beard, J. D., Ball, M. M., Kitras, C., & Lundrigan, P. B. (2026). Long-Term Validation of a Consumer- vs. Industrial-Grade Indoor Radon and Air Quality Monitor at a School in Utah. Sensors, 26(16), 5230. https://doi.org/10.3390/s26165230
Vancouver
Beard JD, Ball MM, Kitras C, Lundrigan PB. Long-Term Validation of a Consumer- vs. Industrial-Grade Indoor Radon and Air Quality Monitor at a School in Utah. Sensors. 2026;26(16):5230. doi:10.3390/s26165230
BibTeX
@article{Beard2026,
author = {Beard, John D and Ball, Maren M and Kitras, Christopher and Lundrigan, Philip B},
title = {Long-Term Validation of a Consumer- vs. Industrial-Grade Indoor Radon and Air Quality Monitor at a School in Utah},
journal = {Sensors},
year = {2026},
volume = {26},
number = {16},
pages = {5230},
doi = {10.3390/s26165230}
}
EUVIMED