Impact of Regeneration on the Dosimetric Properties of LiF:Mg,Cu,P (MCP-N) Detectors

Patrycja Gordon, Katarzyna Matusiak iD, Mariusz Kłosowski iD, Aleksandra Jung

Sensors · 15.08.2026 · 26 (16) · 5176

DOI: 10.3390/s26165176 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 15.08.2026. Verfasst von Patrycja Gordon und 3 weiteren Autor:innen. Beteiligte Einrichtung: Maria Skłodowska-Curie National Research Institute of Oncology, Kracow Branch, 30-059 Krakow, Poland. Die Arbeit nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: AGH University of Krakow, Polish Ministry of Science and Higher Education.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluates the dosimetric properties of standard (SD) and regenerated (RD) MCP-N thermoluminescent detectors over a wide dose range from mGy to Gy, with particular emphasis on the influence of readout mode, dose level, and temporal stability on detector performance. A total of 17 SD and 18 RD detectors were analyzed under standard conditions using two readout modes: a three-step READER mode and a continuous ANALYSER mode. Both detector types exhibit good linearity; however, deviations from ideal behavior at higher doses suggest the involvement of additional trapping and recombination processes. Glow-curve analysis indicates modifications in RD detectors, reflected by enhanced low-temperature peaks and slight shifts in peak position. Temporal effects also influence detector response, with stabilization observed after repeated irradiation–readout cycles. Fading remains low, with relatively stable signal levels over time. The results show that the ANALYSER mode provides improved stability and reproducibility compared to the READER mode, especially at higher doses, where increased variability and certain limitations were observed. Overall, regenerated detectors, despite reduced sensitivity, achieve performance comparable to standard detectors after stabilization. Optimization of readout conditions, particularly the use of ANALYSER mode, is essential for ensuring reliable dosimetric performance across a wide dose range.

Schlagwörter

Stichwörter der Autor:innen

LiF:Mg,Cu,P (MCP-N) detectorsdetector regenerationglow-curve analysisreadout mode reproducibilitythermoluminescent dosimetry

Autor:innen und Institutionen

  • Maria Skłodowska-Curie National Research Institute of Oncology, Kracow Branch, 30-059 Krakow, Poland.
  • AGH University of Krakow, Faculty of Physics and Applied Computer Science, Department of Medical Physics and Biophysics, 30-059 Krakow, Poland.
  • Institute of Nuclear Physics Polish Academy of Sciences, 31-342 Krakow, Poland.

ORCID-Profile: Katarzyna Matusiak, Mariusz Kłosowski

Förderung

  • AGH University of Krakow
  • Polish Ministry of Science and Higher Education

Zitieren

APA

Gordon, P., Matusiak, K., Kłosowski, M., & Jung, A. (2026). Impact of Regeneration on the Dosimetric Properties of LiF:Mg,Cu,P (MCP-N) Detectors. Sensors, 26(16), 5176. https://doi.org/10.3390/s26165176

Vancouver

Gordon P, Matusiak K, Kłosowski M, Jung A. Impact of Regeneration on the Dosimetric Properties of LiF:Mg,Cu,P (MCP-N) Detectors. Sensors. 2026;26(16):5176. doi:10.3390/s26165176

BibTeX
@article{Gordon2026,
  author = {Gordon, Patrycja and Matusiak, Katarzyna and Kłosowski, Mariusz and Jung, Aleksandra},
  title = {Impact of Regeneration on the Dosimetric Properties of LiF:Mg,Cu,P (MCP-N) Detectors},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {16},
  pages = {5176},
  doi = {10.3390/s26165176}
}
Kostenlos starten →