Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide
Sensors · 04.08.2026 · 26 (15) · 4929
DOI: 10.3390/s26154929 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 04.08.2026. Verfasst von Eliana Fernandes und 5 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Pharmacy, University of Coimbra, Health Science Campus, Azinhaga de Santa Comba, 3000-548 Coimbra, Portugal. Die Arbeit nennt 84 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, Foundation for Science and Technology, National Institutes of Health BRAIN Initiative.
Abstract (Originalsprache)
This study investigates the electrochemical properties of polymer-based microelectrode arrays (pMEAs) and their performance in measuring oxygen (O2) and hydrogen peroxide (H2O2). Morphological characterization by scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDS) and X-ray diffraction (XRD) revealed a uniform, fine-grained platinum surface with nanoscale roughness, consistent with the Ti/Pt/Au/Pt multilayer stack architecture. The electrochemical behavior of the pMEAs was assessed using cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS), which demonstrated favorable responses for both O2 reduction and H2O2 oxidation, together with low impedance (41.1 kΩ at 1 kHz). For O2 detection, amperometric measurements at −0.6 V vs. Ag/AgCl indicated a sensitivity of −0.25 ± 0.04 nA μM−1 and a detection limit of 5.4 ± 1.4 nM. For H2O2 detection, application of +0.7 V vs. Ag/AgCl resulted in a sensitivity of 88.13 ± 7.61 nA mM−1 and a detection limit of 41.9 ± 5.6 nM. Selectivity evaluation showed effective interferent exclusion following m-phenylenediamine electrodeposition, without compromising analytical performance. Overall, these findings indicate the suitability of pMEAs for real-time, in vivo monitoring of O2 and H2O2 in brain tissue with high spatial and temporal resolution, supporting applications in oxidative stress research and neurometabolic sensing.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Pharmacy, University of Coimbra, Health Science Campus, Azinhaga de Santa Comba, 3000-548 Coimbra, Portugal.
- CNC-UC-Centre for Neuroscience and Cell Biology, CIBB-Centre for Innovative Biomedicine and Biotechnology, University of Coimbra, Edifício FMUC, Rua Larga, 3004-504 Coimbra, Portugal.
- Alfred E. Mann Department of Biomedical Engineering, University of Southern California, 1042 Downey Way, Denney Research Center (DRB) 140, Los Angeles, CA 90089-1111, USA.
- Centre for Microelectrode Technology (CenMeT), Department of Neuroscience, University of Kentucky Medical Center, MN208, Medical Science Building, 800 Rose Street, Lexington, KY 40536-0298, USA.
- LEC-Laboratory for Electroanalysis and Corrosion, IPN-Instituto Pedro Nunes, Rua Pedro Nunes, 3030-199 Coimbra, Portugal.
ORCID-Profile: Eliana Fernandes, Ana Ledo, Kee Scholten, Greg A Gerhardt, Rui M Barbosa
Förderung
- NIH HHS
- Foundation for Science and Technology
- National Institutes of Health BRAIN Initiative
- European Regional Development Fund
- NIA NIH HHS
- NINDS NIH HHS
- Fundação para a Ciência e Tecnologia
- National Institutes of Health
Zitieren
APA
Fernandes, E., Ledo, A., Scholten, K., Meng, E., Gerhardt, G. A., & Barbosa, R. M. (2026). Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide. Sensors, 26(15), 4929. https://doi.org/10.3390/s26154929
Vancouver
Fernandes E, Ledo A, Scholten K, Meng E, Gerhardt GA, Barbosa RM. Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide. Sensors. 2026;26(15):4929. doi:10.3390/s26154929
BibTeX
@article{Fernandes2026,
author = {Fernandes, Eliana and Ledo, Ana and Scholten, Kee and Meng, Ellis and Gerhardt, Greg A and Barbosa, Rui M},
title = {Electrochemical Evaluation of Polymer-Based Microelectrode Arrays: Analytical Performance on Oxygen and Hydrogen Peroxide},
journal = {Sensors},
year = {2026},
volume = {26},
number = {15},
pages = {4929},
doi = {10.3390/s26154929}
}