Fachartikel in der Zeitschrift Sensors, online erschienen am 25.06.2026. Verfasst von Zhichun Liu und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Aerospace Engineering, Xiamen University, Xiamen 361005, China. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, the Fujian Provincial Natural Science Foundation of China, Fujian Provincial Natural Science Foundation of China.
Abstract (Originalsprache)
High-temperature metallic surfaces in aero-engine hot sections and related thermal systems are subjected to MW/m2-level heat-flux loads and transient thermal conditions, creating a need for sensors capable of quantifying heat flux under high-temperature conditions. This study aims to develop a screen-printed Ag–AgPd thick-film thermopile heat-flux sensor (HFS) for MW/m2-level heat-flux measurement on high-temperature metallic surfaces. Its main feature is the integration of an Ag–AgPd thermopile sensing layer, an insulating layer, and a thermal-resistance layer on a SUS430 stainless-steel substrate through a screen-printing-based multilayer fabrication route. Microstructural characterization, annealing condition comparison, laser comparison calibration, repeated loading, dynamic-response testing, and flame-heating testing were conducted to evaluate the sensor structure and performance. Under laser comparison calibration, the sensor achieved a MW/m2-level calibrated heat-flux response over 0.32–1.37 MW/m2, with a near-linear output relationship of R2>0.998, a sensitivity of 2.67 μV/(kW/m2), a nonlinearity of 1.83%, a hysteresis error below 0.29%, a repeatability error below 0.43%, a sample-to-sample consistency error of 1.06%, a maximum accuracy-test deviation of 1.84%, and a maximum repeated-loading stability error of 1.33%. The sensor also exhibited a time constant of 0.806 s under laser step excitation, and the baseline-corrected equivalent heat-flux response remained stable during approximately 120 s of flame heating at about 800 °C. These results indicate that the proposed HFS provides a feasible thick-film thermopile sensing approach for MW/m2-level heat-flux measurement on high-temperature metallic surfaces.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Aerospace Engineering, Xiamen University, Xiamen 361005, China.
Förderung
- National Natural Science Foundation of China
- the Fujian Provincial Natural Science Foundation of China
- Fujian Provincial Natural Science Foundation of China
Zitieren
APA
Liu, Z., Chen, F., Su, Z., Wang, H., Si, J., Chen, J., Du, Z., & Hai, Z. (2026). Fabrication and Characterization of an Ag–AgPd Thick-Film Thermopile Heat-Flux Sensor for High-Temperature Applications. Sensors, 26(13), 4030. https://doi.org/10.3390/s26134030
Vancouver
Liu Z, Chen F, Su Z, Wang H, Si J, Chen J, et al. Fabrication and Characterization of an Ag–AgPd Thick-Film Thermopile Heat-Flux Sensor for High-Temperature Applications. Sensors. 2026;26(13):4030. doi:10.3390/s26134030
BibTeX
@article{Liu2026,
author = {Liu, Zhichun and Chen, Fei and Su, Zhixuan and Wang, Heng and Si, Jinghan and Chen, Junyang and Du, Zihan and Hai, Zhenyin},
title = {Fabrication and Characterization of an Ag–AgPd Thick-Film Thermopile Heat-Flux Sensor for High-Temperature Applications},
journal = {Sensors},
year = {2026},
volume = {26},
number = {13},
pages = {4030},
doi = {10.3390/s26134030}
}
EUVIMED