Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method

Qingbo Du, Yapeng Yang, Xiaoyu Zhao, Qi Lv, Yuyao Tang, Jiapeng He, Yier Liu, Guoqiang Li

Sensors · 19.06.2026 · 26 (12) · 3913

DOI: 10.3390/s26123913 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 19.06.2026. Verfasst von Qingbo Du und 7 weiteren Autor:innen. Beteiligte Einrichtung: China Institute for Radiation Protection, Taiyuan 030006, China. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The NaI(TI) detector is highly favored in gamma radiation detection and widely applied in fields such as environmental radiation monitoring, nuclear medicine, and laboratory gamma-ray spectroscopy. Its detection performance determines the results of quantitative gamma-ray detection, making it a crucial indicator in detector design and development. This study employs the Monte Carlo method and utilizes TopMC 1.0 software to establish a NaI(TI) detector model. First, the effects of crystal size, ray energy, cladding thickness, and distance on the detector’s detection efficiency were investigated. Subsequently, the energy resolution and peak-to-total ratio of the detector were simulated and calculated, with comparisons made to experimental values. The results indicate that all three detection efficiencies of the NaI(TI) detector are positively correlated with crystal size and exhibit an initial increase followed by a decrease with rising gamma-ray energy. Both the absolute detection efficiency and full-energy peak detection efficiency first increase and then decrease with increasing cladding thickness, while showing a negative correlation with detection distance. The intrinsic detection efficiency is almost unaffected by cladding thickness and initially rises before declining with increasing detection distance. The simulated values of energy resolution closely match experimental values, improving with higher gamma-ray energy. The deviation between simulated and experimental values for different source peak-to-total ratios remains within 6.25%, verifying the model’s reliability and the accuracy of simulation data. These findings provide valuable references and guidance for optimizing detection performance, conducting source-free efficiency calibration, and structural design of NaI(TI) detectors.

Schlagwörter

Stichwörter der Autor:innen

Monte Carlo methodNaI(Tl) detectordetection efficiencydetection performanceenergy resolutionpeak-to-total ratio

Autor:innen und Institutionen

  • China Institute for Radiation Protection, Taiyuan 030006, China.
  • Shanxi Provincial Key Laboratory of Nuclear and Radiation Emergency Response, Taiyuan 030006, China.

Zitieren

APA

Du, Q., Yang, Y., Zhao, X., Lv, Q., Tang, Y., He, J., Liu, Y., & Li, G. (2026). Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method. Sensors, 26(12), 3913. https://doi.org/10.3390/s26123913

Vancouver

Du Q, Yang Y, Zhao X, Lv Q, Tang Y, He J, et al. Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method. Sensors. 2026;26(12):3913. doi:10.3390/s26123913

BibTeX
@article{Du2026,
  author = {Du, Qingbo and Yang, Yapeng and Zhao, Xiaoyu and Lv, Qi and Tang, Yuyao and He, Jiapeng and Liu, Yier and Li, Guoqiang},
  title = {Research on Detection Performance of NaI(Tl) Detector Based on Monte Carlo Method},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {12},
  pages = {3913},
  doi = {10.3390/s26123913}
}
Kostenlos starten →