EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Comparative Evaluation of Machine Learning and Conventional Material Decomposition Algorithms for Spectral Chest Radiography Using a CdTe Photon-Counting Detector

Sriharsha Marupudi, Bahaa Ghammraoui iD

Sensors · 19.05.2026 · 26 (10) · 3202

DOI: 10.3390/s26103202 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 19.05.2026. Verfasst von Sriharsha Marupudi und 1 weiteren Autor:innen. Beteiligte Einrichtung: Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Labs, U.S. Food and Drug Administration, Silver Spring,…. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Spectral chest radiography with photon-counting detectors (PCDs) enables energy-resolved acquisition for bone/soft-tissue separation, but quantitative performance depends on detector cross-talk and the selected material decomposition algorithm. We performed a controlled simulation study comparing a conventional low-order polynomial decomposition model with two machine learning regressors (multilayer perceptron (MLP) and support vector regression (SVR)) for a cadmium telluride (CdTe) PCD. A Geant4-derived detector response model, coupled with a charge-transport model, was integrated into a physics-forward model including charge sharing and Poisson quantum noise. Digital LucAl/IEC 62220-2-1 phantoms with aluminum and polymethyl methacrylate inserts were used for quantitative bias/root mean square error (RMSE) evaluation, and task-based low-contrast detectability that was evaluated using an exponential transformation of the free-response operating characteristic (EFROC) method using a matched-filter template. Performance was evaluated over clinically relevant dose levels (0.07–7.5 mAs), calibration grid densities (3×3 to 8×8), and numbers of energy thresholds (M=2–6). Polynomial decomposition was stable under sparse calibration, whereas ML methods benefited strongly from denser calibration and additional thresholds; SVR achieved the lowest RMSE under dense calibration, while MLP produced smoother maps and improved soft-tissue detectability at low-to-intermediate dose. At high dose, all methods approached near-ideal detection performance. These results quantify practical trade-offs between calibration requirements, quantitative accuracy, and low-contrast detectability for PCD-based spectral chest radiography.

Schlagwörter

Stichwörter der Autor:innen

Monte CarloPoisson statisticschest radiographylow-contrast detectabilitymachine learningmaterial decompositionphoton counting

Autor:innen und Institutionen

  • Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Labs, U.S. Food and Drug Administration, Silver Spring, MD 20993, USA.

ORCID-Profile: Bahaa Ghammraoui

Zitieren

APA

Marupudi, S., & Ghammraoui, B. (2026). Comparative Evaluation of Machine Learning and Conventional Material Decomposition Algorithms for Spectral Chest Radiography Using a CdTe Photon-Counting Detector. Sensors, 26(10), 3202. https://doi.org/10.3390/s26103202

Vancouver

Marupudi S, Ghammraoui B. Comparative Evaluation of Machine Learning and Conventional Material Decomposition Algorithms for Spectral Chest Radiography Using a CdTe Photon-Counting Detector. Sensors. 2026;26(10):3202. doi:10.3390/s26103202

BibTeX
@article{Marupudi2026,
  author = {Marupudi, Sriharsha and Ghammraoui, Bahaa},
  title = {Comparative Evaluation of Machine Learning and Conventional Material Decomposition Algorithms for Spectral Chest Radiography Using a CdTe Photon-Counting Detector},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {10},
  pages = {3202},
  doi = {10.3390/s26103202}
}
Kostenlos starten →