A Device for Characterizing Skin Physiological Response to Mechanical Loading in Transtibial Prosthesis Users

Molly E Baumann iD, Mathew J Weissinger, Joseph L Garbini, Conor L Lanahan, Joseph C Mertens, Bailey Ramesh, W Lee Childers iD, Joan E Sanders iD

Sensors · 08.04.2026 · 26 (8) · 2288

DOI: 10.3390/s26082288 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 08.04.2026. Verfasst von Molly E Baumann und 7 weiteren Autor:innen. Beteiligte Einrichtung: Extremity Trauma and Amputation Center of Excellence, Defense Health Agency, Falls Church, VA 22042, USA. Hauptthemen laut MEDLINE: Artificial Limbs, Skin Physiological Phenomena, Skin, Tibia. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: US Army Medical Research Acquisition Activity, Congressionally Directed Medical Research Program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The skin’s physiological response to repetitive stress is not well understood in prosthesis users. Improving this understanding could facilitate the design of a diagnostic tool to determine if the skin is adapting to tolerate stress from a prosthetic socket. The objective of this research was to develop a physical system that mechanically stresses the skin in a controlled manner and then implements the imaging modalities of infrared (IR) imaging and optical coherence tomography (OCT). IR imaging characterizes the skin’s temperature response, while OCT characterizes vessel diameter changes over time in the skin. The system was implemented in a single individual with a transtibial amputation. The system reliably maintained the force profile throughout testing. IR and OCT imaging were initiated after load application, and all curves demonstrated an initial rise in temperature immediately after load removal followed by a decrease towards baseline. The system was able to effectively detect a peak outcome (temperature and vessel area) with both imaging modalities. The system’s ability to maintain the loading throughout and begin imaging to capture the peak provides promise for expanded use to better understand the skin’s physiological response to loading in prosthesis users. This improved understanding may better inform treatment strategies to optimize patient outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Artificial LimbsSkin Physiological PhenomenaSkinTibiaHumansTomography, Optical CoherenceStress, MechanicalSkin TemperatureWeight-Bearing

Stichwörter der Autor:innen

amputeeangiographycollageninfrared imaginginterface stressoptical coherence tomographyskin adaptationthermal recoveryvasculature

Autor:innen und Institutionen

  • Extremity Trauma and Amputation Center of Excellence, Defense Health Agency, Falls Church, VA 22042, USA.
  • Department of Rehabilitation Medicine, Center for the Intrepid, 3551 Roger Brooke Drive, JBSA Fort Sam Houston, San Antonio, TX 78234, USA.
  • Department of Bioengineering, University of Washington, 355061, 3720 15th Ave NE, Seattle, WA 98195, USA.
  • Department of Mechanical Engineering, University of Washington, 352600, 3720 15th Ave NE, Seattle, WA 98195, USA.

ORCID-Profile: Molly E Baumann, W Lee Childers, Joan E Sanders

Förderung

  • US Army Medical Research Acquisition Activity
  • Congressionally Directed Medical Research Program

Zitieren

APA

Baumann, M. E., Weissinger, M. J., Garbini, J. L., Lanahan, C. L., Mertens, J. C., Ramesh, B., Childers, W. L., & Sanders, J. E. (2026). A Device for Characterizing Skin Physiological Response to Mechanical Loading in Transtibial Prosthesis Users. Sensors, 26(8), 2288. https://doi.org/10.3390/s26082288

Vancouver

Baumann ME, Weissinger MJ, Garbini JL, Lanahan CL, Mertens JC, Ramesh B, et al. A Device for Characterizing Skin Physiological Response to Mechanical Loading in Transtibial Prosthesis Users. Sensors. 2026;26(8):2288. doi:10.3390/s26082288

BibTeX
@article{Baumann2026,
  author = {Baumann, Molly E and Weissinger, Mathew J and Garbini, Joseph L and Lanahan, Conor L and Mertens, Joseph C and Ramesh, Bailey and Childers, W Lee and Sanders, Joan E},
  title = {A Device for Characterizing Skin Physiological Response to Mechanical Loading in Transtibial Prosthesis Users},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {8},
  pages = {2288},
  doi = {10.3390/s26082288}
}
Kostenlos starten →