Tests of HgCdTe Photodetectors Performances for Implementation on the MIST-A Instrument
Sensors · 05.04.2026 · 26 (7) · 2250
DOI: 10.3390/s26072250 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 05.04.2026. Verfasst von Chiara Cencia und 25 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Space Astrophysics and Planetology (IAPS), National Institute of Astrophysics, 00133 Rome, Italy. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Agenzia Spaziale Italiana, Italian Space Agency, National Institute of Astrophysics.
Abstract (Originalsprache)
The Middle-Wave Infrared Imaging Spectrometer for Target Asteroids (MIST-A) will be launched in 2028 aboard the Emirates Mission to the Asteroid belt (EMA) and will operate in the 2–5 μm spectral range to study the asteroids’ surface composition and thermo-physical properties. MIST-A’s Optical Head (OH) design is inherited from the Jovian IR Auroral Mapper (JIRAM), from which the instrument also received two spare Hybrid-Thinned Mercury-Cadmium-Telluride (MCT) photodetectors: the Engineering Model EM2 and the Flight Spare FS1. These are tested to assess their performance after a long period of storage. The laboratory setup for testing both detectors consists of a blackbody and a cryostat which houses the focal plane, maintained at temperatures of 85 K, its nominal operative temperature, and 90 K. Two sets of measurements are performed: (1) characterization of the dark current at different integration times (0 ms, 224 ms, 448 ms, 672 ms, 869 ms, 1120 ms); (2) verification of the detectors’ response linearity, measuring a blackbody at different temperatures (from 50 °C to 100 °C), including ambient temperature (25 °C, with the blackbody turned off). The results of these tests confirm that both models are fully operational and allow us to evaluate the consequences of the years of inactivity on their performance. Through a detailed analysis of the detectors’ properties and a comparison study with the results of the sensors’ first characterization performed by their producer in 2009, we come to the conclusion that both instruments are able to fulfill MIST-A’s scientific requirements. The FS1 displays a better performance with respect to the EM2 and for this has been selected as MIST-A’s Flight Model.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Space Astrophysics and Planetology (IAPS), National Institute of Astrophysics, 00133 Rome, Italy.
- Department of Physics, Tor Vergata University, 00133 Rome, Italy.
- Leonardo Company, 50013 Campi Bisenzio, Italy.
- Italian Space Agency, 00133 Rome, Italy.
ORCID-Profile: Chiara Cencia, Eliana La Francesca, Mauro Ciarniello, Andrea Raponi, Maria Cristina De Sanctis, Simone De Angelis, Michelangelo Formisano, Marco Ferrari, Stefania Stefani, Alessandra Tiberia, Gianrico Filacchione
Förderung
- Agenzia Spaziale Italiana
- Italian Space Agency
- National Institute of Astrophysics
Zitieren
APA
Cencia, C., La Francesca, E., Ciarniello, M., Raponi, A., Capaccioni, F., De Sanctis, M. C., De Angelis, S., Formisano, M., Ferrari, M., Biondi, D., Boccaccini, A., Stefani, S., Piccioni, G., Mura, A., Galiano, A., Tommasi, L., Bartolo, C., Iuzzolino, M., Bucciantini, L., …, & Filacchione, G. (2026). Tests of HgCdTe Photodetectors Performances for Implementation on the MIST-A Instrument. Sensors, 26(7), 2250. https://doi.org/10.3390/s26072250
Vancouver
Cencia C, La Francesca E, Ciarniello M, Raponi A, Capaccioni F, De Sanctis MC, et al. Tests of HgCdTe Photodetectors Performances for Implementation on the MIST-A Instrument. Sensors. 2026;26(7):2250. doi:10.3390/s26072250
BibTeX
@article{Cencia2026,
author = {Cencia, Chiara and La Francesca, Eliana and Ciarniello, Mauro and Raponi, Andrea and Capaccioni, Fabrizio and De Sanctis, Maria Cristina and De Angelis, Simone and Formisano, Michelangelo and Ferrari, Marco and Biondi, David and Boccaccini, Angelo and Stefani, Stefania and Piccioni, Giuseppe and Mura, Alessandro and Galiano, Anna and Tommasi, Leonardo and Bartolo, Clorinda and Iuzzolino, Marcella and Bucciantini, Leda and Dami, Michele and Cossu, Giovanni and Nencioni, Stefano and Olivieri, Angelo and Ammannito, Eleonora and Tiberia, Alessandra and Filacchione, Gianrico},
title = {Tests of HgCdTe Photodetectors Performances for Implementation on the MIST-A Instrument},
journal = {Sensors},
year = {2026},
volume = {26},
number = {7},
pages = {2250},
doi = {10.3390/s26072250}
}
EUVIMED