Near-Infrared Spectroscopy Used During Cardiopulmonary Resuscitation: Instrumentation, Signal Metrics, Clinical Context, and Feasibility: A Scoping Review
Sensors · 30.03.2026 · 26 (7) · 2136
DOI: 10.3390/s26072136 ↗Kurz gefasst
Scoping Review in der Zeitschrift Sensors, online erschienen am 30.03.2026. Verfasst von Zahra Askari und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering, University of British Columbia, Vancouver, BC V6T 1Z4, Canada. Hauptthemen laut MEDLINE: Cardiopulmonary Resuscitation. Die Arbeit nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Reanimation
Abstract (Originalsprache)
Conventional cardiopulmonary resuscitation (CPR) is guided primarily by process metrics that do not directly quantify cerebral hemodynamics or perfusion. Near-infrared spectroscopy (NIRS) provides continuous, non-invasive monitoring of regional tissue oxygenation and has emerged as a candidate modality for physiologic feedback during low-flow states. However, CPR applications vary across devices and signal processing. This scoping review maps how NIRS has been implemented during conventional CPR in humans and porcine models, with emphasis on instrumentation characteristics, signal processing, acquisition bandwidth, artifact handling, physiologic associations, and feasibility constraints. From 1048 records, 39 studies met the inclusion criteria. Most used forehead-based cerebral rSO2 monitoring (30/39). Rising cerebral oxygenation trajectories were consistently associated with return of spontaneous circulation (ROSC). In contrast, persistently low or non-increasing patterns were associated with non-ROSC, and absolute thresholds varied substantially across devices and studies. A minority of investigations derived compression-rate or waveform features from hemoglobin signals. Feasibility findings emphasized rapid probe placement without interrupting compressions but highlighted motion artifact, workflow constraints, and incomplete acquisition reporting. Overall, during conventional CPR, NIRS primarily serves as a dynamic monitor of oxygenation trends rather than a validated prognostic tool. Emerging waveform-based and hemodynamic analyses suggest the potential to evaluate CPR efficiency using perfusion-responsive optical features.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Biomedical Engineering, University of British Columbia, Vancouver, BC V6T 1Z4, Canada.
- International Collaboration on Repair Discoveries, Vancouver, BC V5Z 1M9, Canada.
- Department of Orthopaedics, University of British Columbia, Vancouver, BC V6T 1Z4, Canada.
- Department of Emergency Medicine, University of British Columbia, Vancouver, BC V6T 1Z4, Canada.
ORCID-Profile: Zahra Askari, Mehdi Nourizadeh, Jacob Hutton, Sumaiya Hossain, Calvin Kuo, Jim Christenson, Brian Grunau, Babak Shadgan
Zitieren
APA
Askari, Z., Nourizadeh, M., Hutton, J., Hossain, S., Kuo, C., Christenson, J., Grunau, B., & Shadgan, B. (2026). Near-Infrared Spectroscopy Used During Cardiopulmonary Resuscitation: Instrumentation, Signal Metrics, Clinical Context, and Feasibility: A Scoping Review. Sensors, 26(7), 2136. https://doi.org/10.3390/s26072136
Vancouver
Askari Z, Nourizadeh M, Hutton J, Hossain S, Kuo C, Christenson J, et al. Near-Infrared Spectroscopy Used During Cardiopulmonary Resuscitation: Instrumentation, Signal Metrics, Clinical Context, and Feasibility: A Scoping Review. Sensors. 2026;26(7):2136. doi:10.3390/s26072136
BibTeX
@article{Askari2026,
author = {Askari, Zahra and Nourizadeh, Mehdi and Hutton, Jacob and Hossain, Sumaiya and Kuo, Calvin and Christenson, Jim and Grunau, Brian and Shadgan, Babak},
title = {Near-Infrared Spectroscopy Used During Cardiopulmonary Resuscitation: Instrumentation, Signal Metrics, Clinical Context, and Feasibility: A Scoping Review},
journal = {Sensors},
year = {2026},
volume = {26},
number = {7},
pages = {2136},
doi = {10.3390/s26072136}
}
EUVIMED