Measuring Walking Stability with a Mobile Phone in Older Adults: A Validation Study

Andisheh Bastani iD, Maya G Panisset, L Eduardo Cofré Lizama

Sensors · 25.03.2026 · 26 (7) · 2060

DOI: 10.3390/s26072060 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Sensors, online erschienen am 25.03.2026. Verfasst von Andisheh Bastani und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Allied Health, School of Health Sciences, Swinburne University of Technology, Hawthorn, VIC 3122, Australia. Hauptthemen laut MEDLINE: Cell Phone, Walking. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swinburne University of Technology SoHS Seed Grant, Swinburne University of Technology SoHS Seed Grant https://ror.org/031rekg67.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

(1)

Background

The local divergence exponent (LDE) is a sensitive measure of walking stability deterioration and risk of falling in older adults. We aim to determine the validity the LDE measured using a mobile phone and to assess its ability to discriminate between healthy young and older adults; (2)

Methods

20 older adults (76.4 ± 4.6 years) and 20 young adults (29.1 ± 6.5 yrs) walked for 6 min on a 20-m walkway while wearing a research-grade inertial measurement unit (IMU) and a mobile phone placed on the sternum to record 3D acceleration data. The LDE was calculated using data from both devices for 3D, vertical (VT), mediolateral (ML), anteroposterior (AP), and norm (N) accelerations. ICC (3,1) was used to determine the validity of the mobile phone’s LDE. Mann–Whitney U tests were used to determine age-group discriminability of LDE measures; (3)

Results

LDEs demonstrated excellent absolute agreement between the wearable IMU and mobile phone (ICC = 0.844). Mobile phone-derived LDEs demonstrated excellent validity relative to the wearable IMU (ICC > 0.75). No significant age-related differences in LDE were observed; wearable or mobile sensors (both p > 0.05); (4)

Conclusions

LDEs measures obtained with a mobile phone are valid. No age group differences were identified.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cell PhoneWalkingHumansAgedFemaleMaleAdultAccidental FallsPostural BalanceWearable Electronic DevicesAccelerometry

Stichwörter der Autor:innen

IMULyapunovbalanceelderlyfallssmartphonestability

Autor:innen und Institutionen

  • Department of Allied Health, School of Health Sciences, Swinburne University of Technology, Hawthorn, VIC 3122, Australia.
  • Department of Medicine (Royal Melbourne Hospital), The University of Melbourne, Parkville, VIC 3052, Australia.

ORCID-Profile: Andisheh Bastani

Förderung

  • Swinburne University of Technology SoHS Seed Grant
  • Swinburne University of Technology SoHS Seed Grant https://ror.org/031rekg67

Zitieren

APA

Bastani, A., Panisset, M. G., & Cofré Lizama, L. E. (2026). Measuring Walking Stability with a Mobile Phone in Older Adults: A Validation Study. Sensors, 26(7), 2060. https://doi.org/10.3390/s26072060

Vancouver

Bastani A, Panisset MG, Cofré Lizama LE. Measuring Walking Stability with a Mobile Phone in Older Adults: A Validation Study. Sensors. 2026;26(7):2060. doi:10.3390/s26072060

BibTeX
@article{Bastani2026,
  author = {Bastani, Andisheh and Panisset, Maya G and Cofré Lizama, L Eduardo},
  title = {Measuring Walking Stability with a Mobile Phone in Older Adults: A Validation Study},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {7},
  pages = {2060},
  doi = {10.3390/s26072060}
}