Development of Digital Sampling for Spaceborne Fourier Transform Spectrometers Using Dual Reference Channel
Sensors · 25.03.2026 · 26 (7) · 2036
DOI: 10.3390/s26072036 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 25.03.2026. Verfasst von Andrea Appiani und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Politecnico di Milano, 20156 Milan, Italy. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Agenzia Spaziale Italiana, ASI https://ror.org/034zgem50.
Abstract (Originalsprache)
This work presents an original implementation of the digital sampling pipeline for spaceborne Fourier Transform Spectrometers (FTSs). The implementation aims at improving the robustness of the spectrometer to harsh environmental conditions, including mechanical vibrations and a wide operational temperature range, avoiding the use of dedicated electronic hardware for the interferometer mirrors’ speed control and interferogram sampling. The FTS configuration is based on the constant time step sampling of the interferometer using a standard ADC (Analogue to Digital Converter), along with two metrology laser channels. The development tool is a MATLAB-based simulator developed to emulate the FTS and, in particular, the generation and acquisition of interferograms, incorporating harmonic vibrations and detector noise. The simulator was exploited to compare state-of-the-art techniques and newly implemented variants. An improvement of the arccosine method is first proposed, revising the normalisation process to exploit the full set of recorded data without discarding critical points. Subsequently, methods using two reference channels have been developed and evaluated. Two implementations are considered: two references at the same wavelength with an optimised phase shift (i.e., π/2) and two references at different wavelengths. Different data fusion strategies are compared in terms of spectral uncertainty, varying types of simulated disturbances and noise amplitudes. Results show that the optimal combination of two same-wavelength references consistently outperforms any other configuration, yielding lower average spectral errors and more stable performance over the frequency range and for a lower SNR of reference channels. Conversely, dual-wavelength strategies exhibit reduced accuracy, though they offer flexibility when fixed phase shifts cannot be maintained. The optimal combination of two same-wavelength reference channels, phase-shifted, is a promising configuration for spaceborne FTSs, so the development and test of an instrument breadboard is envisaged as the consequent development of this work.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Politecnico di Milano, 20156 Milan, Italy.
- CISAS-Industrial Engineering Department, Università Degli Studi di Padova, 35131 Padua, Italy.
ORCID-Profile: Andrea Appiani, Diego Scaccabarozzi, Bortolino Saggin
Förderung
- Agenzia Spaziale Italiana
- ASI https://ror.org/034zgem50
Zitieren
APA
Appiani, A., Scaccabarozzi, D., & Saggin, B. (2026). Development of Digital Sampling for Spaceborne Fourier Transform Spectrometers Using Dual Reference Channel. Sensors, 26(7), 2036. https://doi.org/10.3390/s26072036
Vancouver
Appiani A, Scaccabarozzi D, Saggin B. Development of Digital Sampling for Spaceborne Fourier Transform Spectrometers Using Dual Reference Channel. Sensors. 2026;26(7):2036. doi:10.3390/s26072036
BibTeX
@article{Appiani2026,
author = {Appiani, Andrea and Scaccabarozzi, Diego and Saggin, Bortolino},
title = {Development of Digital Sampling for Spaceborne Fourier Transform Spectrometers Using Dual Reference Channel},
journal = {Sensors},
year = {2026},
volume = {26},
number = {7},
pages = {2036},
doi = {10.3390/s26072036}
}