Fachartikel in der Zeitschrift Sensors, online erschienen am 21.03.2026. Verfasst von Christoph Busch und 2 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Technical Medicine (ITeM), Furtwangen University, Jakob-Kienzle-Str. 17, 78054 Villingen-Schwenningen, Germany. Hauptthemen laut MEDLINE: Finite Element Analysis, Electrosurgery. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bundesministerium für Forschung, Technologie und Raumfahrt, German Federal Ministry of Research, Technology and Space.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Physics-based simulations are increasingly used to improve understanding of electrosurgical processes and to enable model-based estimation of tissue state when direct sensing is limited. The performance of such simulation-based virtual sensing approaches strongly depends on an accurate representation of the electrode–tissue interface. Despite its central role in electrical and thermal coupling, the influence of the electrode–tissue contact area has received limited attention in existing simulation studies. In this work, the influence of the electrode–tissue contact area on the sensitivity of key temperature-dependent tissue parameters was investigated for electrosurgical monopolar soft coagulation. Using a multiphysics finite element model under controlled boundary conditions, the sensitivity of maximum temperature development and necrotic tissue volume formation was analyzed with respect to varying contact areas and initial values of electrical conductivity, thermal conductivity, and effective heat capacity. The results demonstrate that parameter sensitivities are strongly contact-area-dependent. Electrical conductivity exhibits the most pronounced influence, particularly at larger contact areas, while thermal conductivity remains of minor relevance. In contrast, effective heat capacity significantly affects necrotic tissue volume formation, with increasing sensitivity for larger contact areas. These findings emphasize the importance of accurately accounting for electrode–tissue contact conditions in simulation-based analyses and clarify how contact-area-dependent sensitivities influence model-based tissue state estimation in electrosurgical coagulation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Finite Element AnalysisElectrosurgeryElectrodesElectric ConductivityComputer SimulationThermal ConductivityHumansTemperature

Stichwörter der Autor:innen

electrode–tissue contact areaelectrosurgical soft coagulationfinite element modelingmultiphysics simulationparameter sensitivity analysis

Autor:innen und Institutionen

  • Institute of Technical Medicine (ITeM), Furtwangen University, Jakob-Kienzle-Str. 17, 78054 Villingen-Schwenningen, Germany.
  • Department of Microsystems Engineering, Faculty of Engineering, University of Freiburg, 79110 Freiburg, Germany.

ORCID-Profile: Christoph Busch, Stefan J Rupitsch, Knut Moeller

Förderung

  • Bundesministerium für Forschung, Technologie und Raumfahrt
  • German Federal Ministry of Research, Technology and Space

Zitieren

APA

Busch, C., Rupitsch, S. J., & Moeller, K. (2026). Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study. Sensors, 26(6), 1975. https://doi.org/10.3390/s26061975

Vancouver

Busch C, Rupitsch SJ, Moeller K. Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study. Sensors. 2026;26(6):1975. doi:10.3390/s26061975

BibTeX
@article{Busch2026,
  author = {Busch, Christoph and Rupitsch, Stefan J and Moeller, Knut},
  title = {Influence of Electrode–Tissue Contact Area on Parameter Sensitivity in Electrosurgical Monopolar Soft Coagulation: A Multiphysics Finite Element Study},
  journal = {Sensors},
  year = {2026},
  volume = {26},
  number = {6},
  pages = {1975},
  doi = {10.3390/s26061975}
}