A Novel Fluorescence-Triggered Auditory Feedback Photosensor for Precision Lymph Node Mapping
Sensors · 10.03.2026 · 26 (6) · 1745
DOI: 10.3390/s26061745 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 10.03.2026. Verfasst von Kicheol Yoon und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Oncology & Gachon Biomedical Convergence Institute, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea. Hauptthemen laut MEDLINE: Lymph Nodes. Die Arbeit nennt 61 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: tech incubator program for startup.
Abstract (Originalsprache)
Background
In cancer surgery, resection of the primary tumor and regional lymph nodes (LNs) is critical. Adequate LN examination is essential to detect metastasis, which determines the cancer stage. Fluorescence emission allows for visual differentiation and rapid monitoring of LNs.
Methods
Cancer tissue is stained with a fluorescent dye (indocyanine green, ICG) to identify LNs. Fluorescence is induced from the stained LNs using LED light, and a photosensor coupled with a speaker detects the fluorescence signal and triggers an audible alarm. Filters are applied to prevent false alarms.
Results
Upon LN detection, an alarm is emitted from the speaker, and the results are recorded using the LED and a digital multimeter (DMM). In clinical trials, ICG is injected to induce LN fluorescence staining, followed by LED irradiation to induce the fluorescent wavelength and verify LN imaging.
Discussion
In clinical trials, ICG stains both LNs and blood vessels, which may lead to false positives. To address this limitation, artificial intelligence algorithms can be trained to specifically identify LNs.
Conclusions
Detection of fluorescence wavelengths via photosensors allows for rapid identification of LNs, confirmed through an audible alarm, thereby reducing surgical time. This method shows potential for broad application in cancer surgery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Division of Medical Oncology & Gachon Biomedical Convergence Institute, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea.
- Medical Devices R&D Center, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea.
- Department of Pre-Medicine, College of Medicine, Gachon University, Seongnam-si 13120, Gyeonggi-do, Republic of Korea.
- Department of Electronic Engineering, Gyeonggi University of Science and Technology, 269 Gyeonggigwagi-dearo, Siheung City 15073, Gyeonggi-do, Republic of Korea.
- Department of Biomedical Engineering, Gachon University, Incheon 21565, Republic of Korea.
- Department of Surgery, Gachon University Gil Medical Center, 21, 774 beon-gil, Namdong-daero, Namdong-gu, Incheon 21565, Republic of Korea.
ORCID-Profile: Hari Kang, Kwang Gi Kim
Förderung
- tech incubator program for startup
Zitieren
APA
Yoon, K., Son, H., Kang, H., Lee, S., Lee, T. H., Lee, W. S., & Kim, K. G. (2026). A Novel Fluorescence-Triggered Auditory Feedback Photosensor for Precision Lymph Node Mapping. Sensors, 26(6), 1745. https://doi.org/10.3390/s26061745
Vancouver
Yoon K, Son H, Kang H, Lee S, Lee TH, Lee WS, et al. A Novel Fluorescence-Triggered Auditory Feedback Photosensor for Precision Lymph Node Mapping. Sensors. 2026;26(6):1745. doi:10.3390/s26061745
BibTeX
@article{Yoon2026,
author = {Yoon, Kicheol and Son, Hyunjun and Kang, Hari and Lee, Sangyun and Lee, Tae-Hyeon and Lee, Won-Suk and Kim, Kwang Gi},
title = {A Novel Fluorescence-Triggered Auditory Feedback Photosensor for Precision Lymph Node Mapping},
journal = {Sensors},
year = {2026},
volume = {26},
number = {6},
pages = {1745},
doi = {10.3390/s26061745}
}
EUVIMED