EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Study on Rejecting Non-Target and Misclassified Motions for Robust Tactile-Sensor-Based Prosthetic Hand Control

Hayato Iwai iD, Feng Wang iD

Sensors · 2026

DOI: 10.3390/s26020721 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2026. Verfasst von Hayato Iwai und 1 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Engineering, Maebashi Institute of Technology, Maebashi 371-0816, Japan. Hauptthemen laut MEDLINE: Hand, Touch, Artificial Limbs. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Reliable motion classification is essential for practical prosthetic-hand control. Unintended activations caused by ambiguous motions, unknown motions, or non-target body movements can degrade controllability and compromise user safety. Mechanical-sensing approaches are attracting attention as alternatives or complements to surface electromyography, and tactile-sensor-based methods represent one such direction. However, despite extensive studies on prosthetic control, systematic investigations of computationally lightweight motion-rejection strategies remain limited. This study investigates rejection mechanisms to improve the robustness of polyvinylidene fluoride (PVDF) tactile-sensor-based prosthetic control. The proposed approach selectively withholds outputs for misclassified and non-target inputs. We compare three mechanisms: (1) one-class support vector machine (OCSVM) outlier detection, (2) entropy-based rejection using a multilayer perceptron (BPNN-Entropy), and (3) a parameter-free decision-consistency check for one-vs-rest support vector machines (SVMs) that withholds classification when the output sign pattern is inconsistent (one-vs-rest reject option (OvR-RO); proposed). Performance is evaluated for three sources of unintended activation: ambiguous target trials (retrospectively defined), unknown motions excluded from training, and non-target body movements. The results show that OvR-RO achieves a favorable balance between rejection rate and rejection precision for ambiguous motions, while maintaining responsiveness. Overall, explicitly rejecting misclassified and non-target motions is effective for enhancing robustness in tactile-sensor-based prosthetic control.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HandTouchArtificial LimbsHumansSupport Vector MachineMotionFluorocarbon PolymersMovementMultilayer PerceptronsPolyvinylsAlgorithmsElectromyography

Stichwörter der Autor:innen

PVDF sensormotion classificationnon-target motionsprosthetic hand controlreject optionrobustnesssupport vector machinetactile sensor

Substanzen

Fluorocarbon Polymerspolyvinylidene fluoridePolyvinyls

Autor:innen und Institutionen

  • Graduate School of Engineering, Maebashi Institute of Technology, Maebashi 371-0816, Japan.
  • Fujitsu Limited, Kawasaki-shi 212-0014, Japan.
  • Faculty of Engineering, Maebashi Institute of Technology, Maebashi 371-0816, Japan.

ORCID-Profile: Hayato Iwai, Feng Wang

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Iwai, H., & Wang, F. (2026). A Study on Rejecting Non-Target and Misclassified Motions for Robust Tactile-Sensor-Based Prosthetic Hand Control. Sensors. https://doi.org/10.3390/s26020721

Vancouver

Iwai H, Wang F. A Study on Rejecting Non-Target and Misclassified Motions for Robust Tactile-Sensor-Based Prosthetic Hand Control. Sensors. 2026. doi:10.3390/s26020721

BibTeX
@article{Iwai2026,
  author = {Iwai, Hayato and Wang, Feng},
  title = {A Study on Rejecting Non-Target and Misclassified Motions for Robust Tactile-Sensor-Based Prosthetic Hand Control},
  journal = {Sensors},
  year = {2026},
  doi = {10.3390/s26020721}
}
Kostenlos starten →