Development of a Robot-Assisted TMS Localization System Using Dual Capacitive Sensors for Coil Tilt Detection
Sensors · 20.01.2026 · 26 (2) · 693
DOI: 10.3390/s26020693 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 20.01.2026. Verfasst von Czaryn Diane Salazar Ompico und 6 weiteren Autor:innen. Beteiligte Einrichtung: Systems and Bioengineering Department, Faculty of Engineering, Maebashi Institute of Technology, Maebashi 371-0816, Gunma, Japan. Hauptthemen laut MEDLINE: Transcranial Magnetic Stimulation, Robotics, Biosensing Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Transcranial Magnetic Stimulation (TMS) is a non-invasive technique for neurological research and therapy, but its effectiveness depends on accurate and stable coil placement. Manual localization based on anatomical landmarks is time-consuming and operator-dependent, while state-of-the-art robotic and neuronavigation systems achieve high accuracy using optical tracking with head-mounted markers and infrared cameras, at the cost of increased system complexity and setup burden. This study presents a cost-effective, markerless robotic-assisted TMS system that combines a 3D depth camera and textile capacitive sensors to assist coil localization and contact control. Facial landmarks detected by the depth camera are used to estimate the motor cortex (C3) location without external tracking markers, while a dual textile-sensor suspension provides compliant “soft-landing” behavior, contact confirmation, and coil-tilt estimation. Experimental evaluation with five participants showed reliable C3 targeting with valid motor evoked potentials (MEPs) obtained in most trials after initial calibration, and tilt-verification experiments revealed that peak MEP amplitudes occurred near balanced sensor readings in 12 of 15 trials (80%). The system employs a collaborative robot designed in accordance with international human–robot interaction safety standards, including force-limited actuation and monitored stopping. These results suggest that the proposed approach can improve the accessibility, safety, and consistency of TMS procedures while avoiding the complexity of conventional optical tracking systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Systems and Bioengineering Department, Faculty of Engineering, Maebashi Institute of Technology, Maebashi 371-0816, Gunma, Japan.
- Evelyn D. Ang-Institute of Biomedical Engineering and Health Technologies, De La Salle University, Manila 1004, Philippines.
- Institute of Brain and Blood Vessels, Mihara Memorial Hospital, Isesaki 372-0006, Gunma, Japan.
- Manufacturing Engineering and Management Department, De La Salle University, Manila 1004, Philippines.
- Gunma Industrial Technology Center, Maebashi 379-2147, Gunma, Japan.
ORCID-Profile: Czaryn Diane Salazar Ompico, Julius Noel Banayo, Masato Odagaki, Yutaka Kikuchi, Armyn Chang Sy, Hirofumi Kurosaki
Zitieren
APA
Ompico, C. D. S., Banayo, J. N., Mashio, Y., Odagaki, M., Kikuchi, Y., Sy, A. C., & Kurosaki, H. (2026). Development of a Robot-Assisted TMS Localization System Using Dual Capacitive Sensors for Coil Tilt Detection. Sensors, 26(2), 693. https://doi.org/10.3390/s26020693
Vancouver
Ompico CDS, Banayo JN, Mashio Y, Odagaki M, Kikuchi Y, Sy AC, et al. Development of a Robot-Assisted TMS Localization System Using Dual Capacitive Sensors for Coil Tilt Detection. Sensors. 2026;26(2):693. doi:10.3390/s26020693
BibTeX
@article{Ompico2026,
author = {Ompico, Czaryn Diane Salazar and Banayo, Julius Noel and Mashio, Yamato and Odagaki, Masato and Kikuchi, Yutaka and Sy, Armyn Chang and Kurosaki, Hirofumi},
title = {Development of a Robot-Assisted TMS Localization System Using Dual Capacitive Sensors for Coil Tilt Detection},
journal = {Sensors},
year = {2026},
volume = {26},
number = {2},
pages = {693},
doi = {10.3390/s26020693}
}
EUVIMED