EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Investigation of a PID-Based Dynamic Illuminance Control System for Intelligent Neonatal Jaundice Phototherapy Using a Blue Light LED Array

Man Xie, Hongjie Zheng, Mei Liu iD, Xing Wen, Yile Fan, Bing-Yuh Lu iD

Sensors · 2026

DOI: 10.3390/s26020528 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2026. Verfasst von Man Xie und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Automation, Guangdong University of Petrochemical Technology, Maoming 525000, China. Hauptthemen laut MEDLINE: Jaundice, Neonatal, Phototherapy, Lighting. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of China, Innovation Project of Guangdong Higher Education, Maoming Green Chemical Industry Research Institute, Guangdong, China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Newborns are unable to reliably express changes in their physical condition due to their physiological immaturity and limited capacity for communication; therefore, continuous and systematic monitoring during phototherapy is essential to ensure timely detection of adverse responses and maintenance of therapeutic safety. This study extends our prior work, which introduced an indirect method for measuring light intensity to improve precision in monitoring newborn skin illumination. Light-emitting diode (LED) phototherapy has attracted considerable attention as an effective treatment for neonatal jaundice (NNJ). This study introduces an three-dimensional configuration of blue LEDs. An Arduino Mega 2560 microcontroller with pulse-width modulation (PWM) technology was employed to independently regulate the intensity of LED strips, enabling precise control of light output. The strips were mounted on an arc-shaped structure that can be adjusted mechanically and electronically through pre-programmed instructions embedded in the microcontroller. The results demonstrate that blue light at a wavelength of 460 ± 10 nm aligns with the peak absorption spectrum of bilirubin, thereby optimizing the efficacy of phototherapy for NNJ. Both observed absorption peaks were within the therapeutically effective range. Computer simulations confirmed that stable output contours can be achieved using rapid electronic scanning with a PID control algorithm to dynamically adjust the duty cycle. Experimental data showed that LED radiation output was largely linear. This supports the use of linear control algorithms and confirms the platform’s feasibility for future research.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Jaundice, NeonatalPhototherapyLightingHumansBlue LightInfant, NewbornAlgorithmsBilirubin

Stichwörter der Autor:innen

Arduino-based controllerblue light LED arrayfeasibility of treatmentincubator design for photo therapy

Substanzen

Bilirubin

Autor:innen und Institutionen

  • School of Automation, Guangdong University of Petrochemical Technology, Maoming 525000, China.
  • School of Engineering, Dali University, Dali 671003, China.
  • College of Mechanical and Electrical Engineering, Shihezi University, Shihezi 832000, China.

ORCID-Profile: Mei Liu, Bing-Yuh Lu

Förderung

  • Natural Science Foundation of China
  • Innovation Project of Guangdong Higher Education
  • Maoming Green Chemical Industry Research Institute, Guangdong, China

Zitieren

APA

Xie, M., Zheng, H., Liu, M., Wen, X., Fan, Y., & Lu, B. Y. (2026). Investigation of a PID-Based Dynamic Illuminance Control System for Intelligent Neonatal Jaundice Phototherapy Using a Blue Light LED Array. Sensors. https://doi.org/10.3390/s26020528

Vancouver

Xie M, Zheng H, Liu M, Wen X, Fan Y, Lu BY. Investigation of a PID-Based Dynamic Illuminance Control System for Intelligent Neonatal Jaundice Phototherapy Using a Blue Light LED Array. Sensors. 2026. doi:10.3390/s26020528

BibTeX
@article{Xie2026,
  author = {Xie, Man and Zheng, Hongjie and Liu, Mei and Wen, Xing and Fan, Yile and Lu, Bing-Yuh},
  title = {Investigation of a PID-Based Dynamic Illuminance Control System for Intelligent Neonatal Jaundice Phototherapy Using a Blue Light LED Array},
  journal = {Sensors},
  year = {2026},
  doi = {10.3390/s26020528}
}