Three-Station Non-Contrast MR Angiography of the Lower Extremities Using Standard and Centric Fresh Blood Imaging
Sensors · 2025
DOI: 10.3390/s25247429 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Won C Bae und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California-San Diego, La Jolla, CA 92093, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Angiography, Lower Extremity, Peripheral Arterial Disease. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Canon Medical Systems (USA), Canon Medical Systems, USA.
Abstract (Originalsprache)
Background
Peripheral artery disease (PAD) is a manifestation of atherosclerosis that affects the extremities, leading to reduced perfusion and functional impairment. Non-contrast magnetic resonance angiography (NC-MRA) provides a safe and quantitative approach for early detection of PAD without the risks associated with contrast agents. The purpose of this study was to demonstrate the application of standard and centric ky-kz FBI techniques for rapid three-station NC-MRA of the entire lower extremity.
Methods
This prospective cross-sectional study compared standard three-station fresh blood imaging (sFBI) with centric ky-kz ordered fresh blood imaging (cFBI) sequences in 10 healthy subjects and 3 patients with PAD (age range: 23–79 years; 7 females) using a 3-Tesla magnetic resonance imaging (MRI) system. Both sequences were acquired at the iliac, femoral, and tibial stations. Image quality (0–4 scale), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were evaluated. Statistical analysis was performed using repeated-measures analysis of variance (ANOVA) with significance set at α = 0.05.
Results
Image quality did not differ significantly between sFBI and cFBI (p = 1.0). The iliac station exhibited lower image quality than the femoral station (p < 0.01). In a PAD patient with an iliac stent, cFBI preserved good image quality in the femoral and tibial stations, whereas sFBI was affected by N/2 aliasing artifacts. Both methods failed to visualize the stented iliac segment. Compared to sFBI, cFBI yielded significantly lower SNR (p < 0.01) and CNR (p < 0.001) but reduced total scan time by approximately 40% (468 s vs. 291 s).
Conclusions
Three-station non-contrast FBI MRA of the peripheral arteries is feasible. The cFBI sequence substantially shortens scan time without compromising diagnostic image quality, offering practical advantages for clinical implementation, improved patient comfort, and reduced motion artifacts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of California-San Diego, La Jolla, CA 92093, USA.
- Department of Radiology, Juntendo University School of Medicine, Tokyo 113-8421, Japan.
- Tobata Kyoritsu Hospital, Kitakyushu 804-0093, Japan.
- Department of Surgery, University of California-San Diego, La Jolla, CA 92037, USA.
ORCID-Profile: Won C Bae, Yoshiki Kuwatsuru, Mitsue Miyazaki
Förderung
- Canon Medical Systems (USA)
- Canon Medical Systems, USA
Zitieren
APA
Bae, W. C., Mesa, A., Malis, V., Kuwatsuru, Y., Nakamura, K., Gaffey, A., & Miyazaki, M. (2025). Three-Station Non-Contrast MR Angiography of the Lower Extremities Using Standard and Centric Fresh Blood Imaging. Sensors. https://doi.org/10.3390/s25247429
Vancouver
Bae WC, Mesa A, Malis V, Kuwatsuru Y, Nakamura K, Gaffey A, et al. Three-Station Non-Contrast MR Angiography of the Lower Extremities Using Standard and Centric Fresh Blood Imaging. Sensors. 2025. doi:10.3390/s25247429
BibTeX
@article{Bae2025,
author = {Bae, Won C and Mesa, Anya and Malis, Vadim and Kuwatsuru, Yoshiki and Nakamura, Katsumi and Gaffey, Ann and Miyazaki, Mitsue},
title = {Three-Station Non-Contrast MR Angiography of the Lower Extremities Using Standard and Centric Fresh Blood Imaging},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25247429}
}
EUVIMED