MoveLab®: Validation and Development of Novel Cross-Platform Gait and Mobility Assessments Using Gold Standard Motion Capture and Clinical Standard Assessment
Sensors · 12.09.2025 · 25 (18) · 5706
DOI: 10.3390/s25185706 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 12.09.2025. Verfasst von Katie Powell und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Engineering, Cardiff University, Cardiff CF10 3AT, UK. Hauptthemen laut MEDLINE: Gait, Gait Analysis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovate UK, Innovate UK Advancing Precision Medicine 2023–2025, Accelerate Wales, Life Sciences Hub Programme from Welsh Government European Regional Development.
Themen: Mobilisation
Abstract (Originalsprache)
Wearable health assessment devices enable real-time clinical- and home-based patient monitoring. Human gait analysis is a widely accepted musculoskeletal assessment. The 30 s Sit-to-Stand (STS) and Timed-Up-and-Go (TUG) are clinical frailty assessments used alongside gait analysis. This study assessed the reliability and validity of the MoveLab® (Agile Kinetic 2024) approach to measure gait spatiotemporal parameters (STPs), STS, and TUG using a waist-worn mobile phone, compared to the Gold Standard 3D marker-based motion capture (Qualisys AB, Sweden) and the Clinical Standard assessment of the STS and TUG test methods. Movement data, recorded simultaneously for 25 healthy volunteers (14 female and 11 male, Age = 31.8 ± 11.6 yrs) in a Biomechanics Laboratory using the Gold Standard system, the Clinical Standard assessments, and MoveLab®, was analyzed using Intraclass Correlation (ICC) and Bland–Altman plots (Python) to quantify the correlations, consistency, and significance across the output parameters. Comparing the methods, the STP consistency ranged from acceptable to good for all the tested parameters (ICC 0.299–0.894). The highest and lowest correlations were cycle time and terminal double support time, respectively. The TUG showed good agreement (ICC 0.757). Generally, an equal number of MoveLab® STS repetitions were observed. MoveLab® demonstrated validity and reliability for a range of key movement parameters using a pouch-worn mobile phone device in healthy adults in a controlled laboratory environment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Engineering, Cardiff University, Cardiff CF10 3AT, UK.
- Agile Kinetic, Newport NP20 1GL, UK.
ORCID-Profile: Ahmad Amer, Zornitza Glavcheva-Laleva, Jenny Williams, Peter Bishop, Catherine Holt
Förderung
- Innovate UK
- Innovate UK Advancing Precision Medicine 2023–2025
- Accelerate Wales, Life Sciences Hub Programme from Welsh Government European Regional Development
Zitieren
APA
Powell, K., Amer, A., Glavcheva-Laleva, Z., Williams, J., O'Flaherty Farrell, C., Harwood, F., Bishop, P., & Holt, C. (2025). MoveLab®: Validation and Development of Novel Cross-Platform Gait and Mobility Assessments Using Gold Standard Motion Capture and Clinical Standard Assessment. Sensors, 25(18), 5706. https://doi.org/10.3390/s25185706
Vancouver
Powell K, Amer A, Glavcheva-Laleva Z, Williams J, O'Flaherty Farrell C, Harwood F, et al. MoveLab®: Validation and Development of Novel Cross-Platform Gait and Mobility Assessments Using Gold Standard Motion Capture and Clinical Standard Assessment. Sensors. 2025;25(18):5706. doi:10.3390/s25185706
BibTeX
@article{Powell2025,
author = {Powell, Katie and Amer, Ahmad and Glavcheva-Laleva, Zornitza and Williams, Jenny and O'Flaherty Farrell, Caomhnad and Harwood, Finchley and Bishop, Peter and Holt, Catherine},
title = {MoveLab®: Validation and Development of Novel Cross-Platform Gait and Mobility Assessments Using Gold Standard Motion Capture and Clinical Standard Assessment},
journal = {Sensors},
year = {2025},
volume = {25},
number = {18},
pages = {5706},
doi = {10.3390/s25185706}
}
EUVIMED