Estimating Heart Rate from Inertial Sensors Embedded in Smart Eyewear: A Validation Study
Sensors · 2025
DOI: 10.3390/s25154531 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Sensors, 2025. Verfasst von Sarah Solbiati und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics, Information and Bioengineering, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milan, MI, Italy. Hauptthemen laut MEDLINE: Heart Rate, Smart Glasses. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Smart glasses are promising alternatives for the continuous, unobtrusive monitoring of heart rate (HR). This study validates HR estimates obtained with the “Essilor Connected Glasses” (SmartEW) during sedentary activities. Thirty participants wore the SmartEW, equipped with an IMU sensor for HR estimation, a commercial smartwatch (Garmin Venu 3), and an ECG device (Movesense Flash). The protocol included six static tasks performed under controlled laboratory conditions. The SmartEW algorithm analyzed 22.5 s signal windows using spectral analysis to estimate HR and provide a quality index (QI). Statistical analyses assessed agreement with ECG and the impact of QI on HR accuracy. SmartEW showed high agreement with ECG, especially with QI threshold equal to 70, as a trade-off between accuracy, low error, and acceptable data coverage (80%). Correlation for QI ≥ 70 was high across all the experimental phases (r2 up to 0.96), and the accuracy within ±5 bpm reached 95%. QI ≥ 70 also allowed biases to decrease (e.g., from −1.83 to −0.19 bpm while standing), with narrower limits of agreement, compared to ECG. SmartEW showed promising HR accuracy across sedentary activities, yielding high correlation and strong agreement with ECG and Garmin. SmartEW appears suitable for HR monitoring in static conditions, particularly when data quality is ensured.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electronics, Information and Bioengineering, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milan, MI, Italy.
- EssilorLuxottica, Smart Eyewear Lenses, Essilor International, 147 Rue de Paris, 94320 Créteil, France.
- EssilorLuxottica, Smart Eyewear Lab, Piazzale Luigi Cadorna 3, 20123 Milan, MI, Italy.
- IRCCS Istituto Auxologico Italiano, Via Ludovico Ariosto 13, 20145 Milan, MI, Italy.
ORCID-Profile: Sarah Solbiati, Federica Mozzini, Niccolò Antonello, Diana Trojaniello, Enrico Gianluca Caiani
Zitieren
APA
Solbiati, S., Mozzini, F., Sahler, J., Gil, P., Amir, B., Antonello, N., Trojaniello, D., & Caiani, E. G. (2025). Estimating Heart Rate from Inertial Sensors Embedded in Smart Eyewear: A Validation Study. Sensors. https://doi.org/10.3390/s25154531
Vancouver
Solbiati S, Mozzini F, Sahler J, Gil P, Amir B, Antonello N, et al. Estimating Heart Rate from Inertial Sensors Embedded in Smart Eyewear: A Validation Study. Sensors. 2025. doi:10.3390/s25154531
BibTeX
@article{Solbiati2025,
author = {Solbiati, Sarah and Mozzini, Federica and Sahler, Jean and Gil, Paul and Amir, Bruno and Antonello, Niccolò and Trojaniello, Diana and Caiani, Enrico Gianluca},
title = {Estimating Heart Rate from Inertial Sensors Embedded in Smart Eyewear: A Validation Study},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25154531}
}