A Line-Source Approach for Simulating MammoWave Microwave Imaging Apparatus for Breast Lesion Detection
Sensors · 2025
DOI: 10.3390/s25123640 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Navid Ghavami und 3 weiteren Autor:innen. Beteiligte Einrichtung: UBT-Umbria Bioengineering Technologies, 06081 Perugia, Italy. Hauptthemen laut MEDLINE: Breast Neoplasms, Microwave Imaging, Microwaves. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union's Horizon research and innovation framework programme, MammoScreen project.
Abstract (Originalsprache)
Here, we propose an analytical approach to simulating MammoWave, a novel apparatus for breast cancer detection using microwave imaging. The approach is built upon the theory of cylindrical waves emitted by line sources. The sample is modelled as a cylinder with an inclusion. Our results indicate that when compared with phantom measurements, our approach gives an average relative error (between the image generated through measurement with phantoms and the image generated through the analytical simulation approach) of less than 6% when considering the full frequency band of 1–9 GHz. The procedure permits the simulation of the MammoWave imaging system loaded with multilayered eccentric cylinders; thus, it can be used to obtain an insight into MammoWave’s detection capability, without having to perform either time-consuming full-wave simulations or phantom measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- UBT-Umbria Bioengineering Technologies, 06081 Perugia, Italy.
- School of Engineering, London South Bank University, London SE1 0AA, UK.
ORCID-Profile: Navid Ghavami, Sandra Dudley, Mohammad Ghavami, Gianluigi Tiberi
Förderung
- European Union's Horizon research and innovation framework programme
- MammoScreen project
Zitieren
APA
Ghavami, N., Dudley, S., Ghavami, M., & Tiberi, G. (2025). A Line-Source Approach for Simulating MammoWave Microwave Imaging Apparatus for Breast Lesion Detection. Sensors. https://doi.org/10.3390/s25123640
Vancouver
Ghavami N, Dudley S, Ghavami M, Tiberi G. A Line-Source Approach for Simulating MammoWave Microwave Imaging Apparatus for Breast Lesion Detection. Sensors. 2025. doi:10.3390/s25123640
BibTeX
@article{Ghavami2025,
author = {Ghavami, Navid and Dudley, Sandra and Ghavami, Mohammad and Tiberi, Gianluigi},
title = {A Line-Source Approach for Simulating MammoWave Microwave Imaging Apparatus for Breast Lesion Detection},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25123640}
}