Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 07.04.2025. Verfasst von Taeheon Kim und 1 weiteren Autor:innen. Beteiligte Einrichtung: School of Electrical Engineering, Korea University, Seoul 136-713, Republic of Korea. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation, Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT and Future Planning.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We demonstrate a flexible electrochemical biosensor for non-enzymatic glucose detection under different bending conditions. The novel flexible glucose sensor consists of a Cu nanoparticle (NP)/laser-induced graphene fiber (LIGF)/porous laser-induced graphene (LIG) network structure on a polyimide film. The bare LIGF/LIG electrode fabricated using an 8.9 W laser power shows a measured sheet resistance and thickness of 6.8 Ω/□ and ~420 μm, respectively. In addition, a conventional Cu NP electroplating method is used to fabricate a Cu/LIGF/LIG electrode-based glucose sensor that shows excellent glucose detection characteristics, including a sensitivity of 1438.8 µA/mM∙cm2, a limit of detection (LOD) of 124 nM, and a broad linear range at an applied potential of +600 mV. Significantly, the Cu/LIGF/LIG electrode-based glucose sensor exhibits a relatively high sensitivity, low LOD, good linear detection range, and long-term stability at bending angles of 0°, 45°, 90°, 135°, and 180°.

Schlagwörter

Stichwörter der Autor:innen

Cu nanoparticleflexible sensorlaser-induced graphenenon-enzymatic electrochemical detection

Autor:innen und Institutionen

  • School of Electrical Engineering, Korea University, Seoul 136-713, Republic of Korea.

ORCID-Profile: Taeheon Kim, James Jungho Pak

Förderung

  • National Research Foundation
  • Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT and Future Planning

Zitieren

APA

Kim, T., & Pak, J. J. (2025). Fabrication and Characterization of a Flexible Non-Enzymatic Electrochemical Glucose Sensor Using a Cu Nanoparticle/Laser-Induced Graphene Fiber/Porous Laser-Induced Graphene Network Electrode. Sensors, 25(7), 2341. https://doi.org/10.3390/s25072341

Vancouver

Kim T, Pak JJ. Fabrication and Characterization of a Flexible Non-Enzymatic Electrochemical Glucose Sensor Using a Cu Nanoparticle/Laser-Induced Graphene Fiber/Porous Laser-Induced Graphene Network Electrode. Sensors. 2025;25(7):2341. doi:10.3390/s25072341

BibTeX
@article{Kim2025,
  author = {Kim, Taeheon and Pak, James Jungho},
  title = {Fabrication and Characterization of a Flexible Non-Enzymatic Electrochemical Glucose Sensor Using a Cu Nanoparticle/Laser-Induced Graphene Fiber/Porous Laser-Induced Graphene Network Electrode},
  journal = {Sensors},
  year = {2025},
  volume = {25},
  number = {7},
  pages = {2341},
  doi = {10.3390/s25072341}
}
Kostenlos starten →