Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Murugaiya Sridar Ilango und 5 weiteren Autor:innen. Beteiligte Einrichtung: Nanobioelectronics Laboratory, Department of Biomedical Engineering, Ben-Gurion University of the Negev, Beer Sheva 8410501, Israel. Hauptthemen laut MEDLINE: Olanzapine, Antipsychotic Agents, Platinum, Cell Membrane Permeability. Die Arbeit nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Israel Science Foundation.
Abstract (Originalsprache)
The blood–brain barrier (BBB) is key to the regular functioning of the central nervous system. The dysfunction of the BBB has been described in various neurological disorders, including schizophrenia. Schizophrenia (SCZ) is a chronic psychiatric disorder described by hallucinations, delusions, and negative symptoms. The Olanzapine (OLZ) drug is an electroactive species, and its levels can be monitored using electrochemical sensors. The detection of OLZ was demonstrated previously by using electrochemical sensors, and this technique can be used to monitor the levels of OLZ in real time. The challenge is to identify the permeability of OLZ through the BBB, so a replica model was designed with the BBB based on a Transwell membrane seeded with endothelial cells. A microfabricated electrode consisting of a 3 mm Au disk was modified with platinum black; this enables higher selectivity of electrochemical signals from OLZ. The dose–response of OLZ was characterized in phosphate buffer saline solution (10 mM, pH 7.4) by adding 20–200 nM (in steps 20) of OLZ stock solution. The observed chronoamperometric electrochemical signals showed an increasing current at 0.45 V vs. Ag/AgCl with an increasing OLZ concentration. The controls for the experiments were performed in phosphate-buffered saline solution (10 mM, pH 7.4). The detection limit was calculated as 9.96 ± 7.35 × 10−6 nM from the calibration curve. The membrane permeability of the OLZ drug tested with five SCZ patients was monitored by studying the TEER measurements and permeability rate constant data.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Nanobioelectronics Laboratory, Department of Biomedical Engineering, Ben-Gurion University of the Negev, Beer Sheva 8410501, Israel.
- Department of Physiology and Cell Biology, Faculty of Health Sciences, Ben-Gurion University of the Negev, Beer Sheva 8410501, Israel.
ORCID-Profile: Murugaiya Sridar Ilango, Gad Vatine, Hadar Ben-Yoav
Förderung
- Israel Science Foundation
Zitieren
APA
Ilango, M. S., Desagani, D., Jagadeesan, S., Snezhko, A., Vatine, G., & Ben-Yoav, H. (2025). Membrane Permeability Monitoring to Antipsychotic Olanzapine Using Platinum Black-Modified Electrodes. Sensors. https://doi.org/10.3390/s25072266
Vancouver
Ilango MS, Desagani D, Jagadeesan S, Snezhko A, Vatine G, Ben-Yoav H. Membrane Permeability Monitoring to Antipsychotic Olanzapine Using Platinum Black-Modified Electrodes. Sensors. 2025. doi:10.3390/s25072266
BibTeX
@article{Ilango2025,
author = {Ilango, Murugaiya Sridar and Desagani, Dayananda and Jagadeesan, Srikanth and Snezhko, Alexander and Vatine, Gad and Ben-Yoav, Hadar},
title = {Membrane Permeability Monitoring to Antipsychotic Olanzapine Using Platinum Black-Modified Electrodes},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25072266}
}
EUVIMED