EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor

Ariel B Kauffman iD, Ruben Shakya, Shuai Yu, Mark A Arnold iD

Sensors · 2025

DOI: 10.3390/s25072238 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Ariel B Kauffman und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering, Rockley Photonics, Irvine, CA 92612, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Glucose, Biosensing Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Benchmark data are reported for a solid-state laser-based near-infrared spectrometer designed for noninvasive measurements in human skin. These data were obtained using a set of aqueous phantoms composed of polystyrene beads, triton X-100, saline, and glucose. The performance of this prototype solid-state laser platform was compared to parallel results obtained with a Fourier-transform (FT) spectrometer. The fundamental spectroscopic performances of the two spectrometer systems were quantified by an analysis of 100% lines determined by ratioing back-to-back spectra collected over time for each phantom. Root mean square (RMS) noise levels were computed for each dataset and the median RMS noise levels were 327.8 µAU and 667.2 µAU for the FT spectrometer and prototype laser platform, respectively. The analytical utility of the solid-state laser platform was assessed through a series of leave-one-phantom-out partial least squares analyses. Results for the laser prototype data included a standard error of cross validation (SECV) of 7.82 mg/dL for an optimized PLS model with 10 factors over a spectral range of 1401–2238 nm. This compares favorably with the results from the FT spectrometer of an SECV of 6.62 mg/dL with 8 factors and a spectral range of 1551–2378 nm. The additional two PLS factors for the laser prototype were shown to be a consequence of its higher spectral noise. Selectivity of these PLS models was demonstrated by comparing models associated with correct and random glucose assignments to each spectrum. Overall, these findings benchmark the analytical utility of this solid-state laser prototype.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingGlucoseBiosensing TechniquesHumansSpectroscopy, Near-InfraredSkinLeast-Squares AnalysisSpectroscopy, Fourier Transform Infrared

Stichwörter der Autor:innen

RMS spectral noise benchmarkingnear-infrared spectroscopynoninvasive glucose monitornoninvasive glucose sensingphotonic integrated chipsskin glucose phantomssolid-state laser spectroscopy

Substanzen

Glucose

Autor:innen und Institutionen

  • Department of Engineering, Rockley Photonics, Irvine, CA 92612, USA.
  • Department of Chemistry, University of Iowa, Iowa City, IA 52242, USA.

ORCID-Profile: Ariel B Kauffman, Mark A Arnold

Zitieren

APA

Kauffman, A. B., Shakya, R., Yu, S., & Arnold, M. A. (2025). Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor. Sensors. https://doi.org/10.3390/s25072238

Vancouver

Kauffman AB, Shakya R, Yu S, Arnold MA. Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor. Sensors. 2025. doi:10.3390/s25072238

BibTeX
@article{Kauffman2025,
  author = {Kauffman, Ariel B and Shakya, Ruben and Yu, Shuai and Arnold, Mark A},
  title = {Noninvasive Glucose Measurements in Tissue Simulating Phantoms Using a Solid-State Near-Infrared Sensor},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25072238}
}
Kostenlos starten →