EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

An IoT-Enabled Wearable Device for Fetal Movement Detection Using Accelerometer and Gyroscope Sensors

Atcharawan Rattanasak iD, Talit Jumphoo, Wongsathon Pathonsuwan, Kasidit Kokkhunthod iD, Khwanjit Orkweha, Khomdet Phapatanaburi, Pattama Tongdee, Porntip Nimkuntod, Monthippa Uthansakul, Peerapong Uthansakul iD

Sensors · 2025

DOI: 10.3390/s25051552 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Atcharawan Rattanasak und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Telecommunication Engineering, Suranaree University of Technology, Nakhon Ratchasima 30000, Thailand. Hauptthemen laut MEDLINE: Wearable Electronic Devices, Fetal Movement, Accelerometry, Fetal Monitoring. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 68 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Council of Thailand (NRCT), Suranaree University of Technology (SUT), Thailand Science Research and Innovation (TSRI).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Counting fetal movements is essential for assessing fetal health, but manually recording these movements can be challenging and inconvenient for pregnant women. This study presents a wearable device designed to detect fetal movements across various settings, both within and outside medical facilities. The device integrates accelerometer and gyroscope sensors with Internet of Things (IoT) technology to accurately differentiate between fetal and non-fetal movements. Data were collected from 35 pregnant women at Suranaree University of Technology (SUT) Hospital. This study evaluated ten signal extraction methods, six machine learning algorithms, and four feature selection techniques to enhance classification performance. The device utilized Particle Swarm Optimization (PSO) for feature selection and Extreme Gradient Boosting (XGB) with PSO hyper-tuning. It achieved a sensitivity of 90.00%, precision of 87.46%, and an F1-score of 88.56%, reflecting commendable results. The IoT-enabled technology facilitated continuous monitoring with an average latency of 423.6 ms. It ensured complete data integrity and successful transmission, with the capability to operate continuously for up to 48 h on a single charge. The findings substantiate the efficacy of the proposed approach in detecting fetal movements, thereby demonstrating a practical and valuable technology for fetal movement detection applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Wearable Electronic DevicesFetal MovementAccelerometryFetal MonitoringHumansFemalePregnancyAlgorithmsInternet of ThingsMachine LearningSignal Processing, Computer-AssistedAdult

Stichwörter der Autor:innen

fetal movement detectioninternet of thingsmachine learningwearable device

Autor:innen und Institutionen

  • School of Telecommunication Engineering, Suranaree University of Technology, Nakhon Ratchasima 30000, Thailand.
  • Institute of Research and Development, Suranaree University of Technology, Nakhon Ratchasima 30000, Thailand.
  • Department of Integrated Engineering, Rajamangala University of Technology Tawan-Ok, Chanthaburi 22210, Thailand.
  • Department of Telecommunication Engineering, Faculty of Engineering and Technology, Rajamangala University of Technology Isan (RMUTI), Nakhon Ratchasima 30000, Thailand.
  • School of Obstetrics and Gynecology, Institute of Medicine, Suranaree University of Technology, Nakhon Ratchasima 30000, Thailand.
  • School of Medicine, Institute of Medicine, Suranaree University of Technology, Nakhon Ratchasima 30000, Thailand.

ORCID-Profile: Atcharawan Rattanasak, Kasidit Kokkhunthod, Peerapong Uthansakul

Förderung

  • National Research Council of Thailand (NRCT)
  • Suranaree University of Technology (SUT)
  • Thailand Science Research and Innovation (TSRI)
  • National Science, Research and Innovation Fund (NSRF) through NRIIS

Zitieren

APA

Rattanasak, A., Jumphoo, T., Pathonsuwan, W., Kokkhunthod, K., Orkweha, K., Phapatanaburi, K., Tongdee, P., Nimkuntod, P., Uthansakul, M., & Uthansakul, P. (2025). An IoT-Enabled Wearable Device for Fetal Movement Detection Using Accelerometer and Gyroscope Sensors. Sensors. https://doi.org/10.3390/s25051552

Vancouver

Rattanasak A, Jumphoo T, Pathonsuwan W, Kokkhunthod K, Orkweha K, Phapatanaburi K, et al. An IoT-Enabled Wearable Device for Fetal Movement Detection Using Accelerometer and Gyroscope Sensors. Sensors. 2025. doi:10.3390/s25051552

BibTeX
@article{Rattanasak2025,
  author = {Rattanasak, Atcharawan and Jumphoo, Talit and Pathonsuwan, Wongsathon and Kokkhunthod, Kasidit and Orkweha, Khwanjit and Phapatanaburi, Khomdet and Tongdee, Pattama and Nimkuntod, Porntip and Uthansakul, Monthippa and Uthansakul, Peerapong},
  title = {An IoT-Enabled Wearable Device for Fetal Movement Detection Using Accelerometer and Gyroscope Sensors},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25051552}
}
Kostenlos starten →