Pressure Sensors for Measuring Tibiofemoral Contact Mechanics in Meniscal Root Repair: A Systematic Review
Sensors · 2025
DOI: 10.3390/s25051507 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Sensors, 2025. Verfasst von Khalis Boksh und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Birmingham, Birmingham B15 2TT, UK. Hauptthemen laut MEDLINE: Tibia, Femur, Menisci, Tibial. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 67 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Orthopaedic Research UK, Royal College of Surgeons of Edinburgh, joint ORUK-RCSEd 2023 research fellowship.
Abstract (Originalsprache)
Background
Tibiofemoral contact mechanics (TFCM) is an accepted biomechanical metrics for evaluating the meniscus in its intact, torn, and repaired states. Pressure sensors are increasingly used, with accuracy and repeatability influenced by test conditions, their design, and their properties. To identify factors optimising performance, we performed a systematic review of the literature on their use for measuring TFCM in posterior meniscal root tears.
Methods
The Cochrane Controlled Register of Trials, PubMed, and Embase were used to perform a systematic review using the PRISMA criteria. As laboratory and surgical setup can influence sensor performance, we collected data on specimen preparation, repair techniques, hardware use, and biomechanical testing parameters.
Results
24 biomechanical studies were included. Specimen preparations were similar across studies with respect to femoral and tibial mounting. Single axial compressive forces were applied between 100 and 1800 N at varying flexion angles (0–90°). Tekscan (Boston, MA, USA) was the commonest sensor used to measure TFCM, followed by digital capacitive sensors and Fujifilm (Tokyo, Japan). Factors influencing their performance included fluid exposure, lack of adequate fixation, non-specific calibration protocols, load saturation exceeding calibration, damaged sensels and inappropriate pre-test conditioning.
Conclusions
Understanding potential factors influencing pressure sensors may improve accuracy, area, and pressure distribution measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of Birmingham, Birmingham B15 2TT, UK.
- Leicester Academic Knee Unit, University Hospitals of Leicester NHS Trust, Leicester LE1 5WW, UK.
- Department of Trauma and Orthopaedics, Morriston Hospital, Swansea SA6 6NL, UK.
- Department of Trauma & Orthopaedics, University Hospitals of Birmingham NHS Trust, Birmingham B15 2GW, UK.
ORCID-Profile: Duncan E T Shepherd, Daniel M Espino, Randeep Aujla
Förderung
- Orthopaedic Research UK
- Royal College of Surgeons of Edinburgh
- joint ORUK-RCSEd 2023 research fellowship
Zitieren
APA
Boksh, K., Bashabayev, B., Shepherd, D. E. T., Espino, D. M., Ghosh, A., Aujla, R., & Boutefnouchet, T. (2025). Pressure Sensors for Measuring Tibiofemoral Contact Mechanics in Meniscal Root Repair: A Systematic Review. Sensors. https://doi.org/10.3390/s25051507
Vancouver
Boksh K, Bashabayev B, Shepherd DET, Espino DM, Ghosh A, Aujla R, et al. Pressure Sensors for Measuring Tibiofemoral Contact Mechanics in Meniscal Root Repair: A Systematic Review. Sensors. 2025. doi:10.3390/s25051507
BibTeX
@article{Boksh2025,
author = {Boksh, Khalis and Bashabayev, Beibit and Shepherd, Duncan E T and Espino, Daniel M and Ghosh, Arijit and Aujla, Randeep and Boutefnouchet, Tarek},
title = {Pressure Sensors for Measuring Tibiofemoral Contact Mechanics in Meniscal Root Repair: A Systematic Review},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25051507}
}
EUVIMED