Evaluation of Pulse Oximetry Accuracy in a Commercial Smartphone and Smartwatch Device During Human Hypoxia Laboratory Testing

Sara H Browne, Michael Bernstein, Philip E Bickler

Sensors · 2025

DOI: 10.3390/s25051286 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Sara H Browne und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, University of California San Diego, La Jolla, CA 92093, USA. Hauptthemen laut MEDLINE: Oximetry, Smartphone, Hypoxia. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIMH NIH HHS, NIH HHS, NIH.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The US Food and Drug Administration (FDA) and International Organization for Standardization (ISO) clearance standards for the clinical use of smart device pulse oximetry require in-laboratory human hypoxemia testing in healthy human individuals using arterial blood gas analysis.

Methods

We evaluated the SpO2 measurements of the Samsung smartphone (Galaxy S9/10) and smartwatch (Galaxy 4) at stable arterial oxygen saturations (SaO2) between 70 and 100% in 24 healthy participants. Testing followed FDA/ISO-stipulated procedures for pulse oximetry performance validation, which include questionnaire estimation of skin tone based on Fitzpatrick estimation of skin types I–VI. During testing, inspired oxygen, nitrogen, and carbon dioxide partial pressures were monitored and adjusted via partial rebreathing circuits to achieve stable target arterial blood oxygen (SaO2) plateaus between 70% and 100%. Arterial blood samples were taken at each plateau, with device SpO2 readings taken at each sample extraction. An ABL-90FLEX blood gas analyzer determined arterial blood sample SaO2. Bias, calculated from device readings minus corresponding arterial blood measurements, was reported as root mean square deviation (RMSD).

Results

Combined Participants demographics were: 62.5% female; median age 26 years (range 21–46); and race/ethnicity 16.7% African American, 33.3% Asian, 12.5% multi-ethnic, and 37.5% Caucasian. Fitzpatrick Skin Scale-identified skin tones were: white–fair (I&II), 20.8%; average–light brown (III–IV), 54% and brown–black (V–VI), 25%. There were no adverse events. The RMSD values of SpO2 measurements were: smartphone 2.6% (257 data pairs) and smartwatch 1.8% (247 data pairs).

Conclusions

Device SpO2 demonstrated RMSD < 3.0% to SaO2, meeting FDA/ISO clearance standards at the time of study. However, additional testing in persons with darker skin tones is necessary. Smartphones and paired wearables, when cleared for clinical use following revision of FDA clearance standards, may expand access to remote pulse oximetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetrySmartphoneHypoxiaHumansFemaleAdultMaleMiddle AgedOxygenYoung AdultOxygen SaturationBlood Gas Analysis

Stichwörter der Autor:innen

FDA clearancehealthhuman hypoxia testingpulse oximetrysmartphonesmartphone sensorssmartphone-paired wearables

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Medicine, University of California San Diego, La Jolla, CA 92093, USA.
  • Specialists in Global Health, Encinitas, CA 92024, USA.
  • Physio Monitor, LLC, San Ramon, CA 94583, USA.
  • Department of Anesthesia, University of California San Francisco, San Francisco, CA 94143, USA.

Förderung

  • NIMH NIH HHS
  • NIH HHS
  • NIH
  • on-profit Specialists in Global Health
  • Samsung Electronics

Zitieren

APA

Browne, S. H., Bernstein, M., & Bickler, P. E. (2025). Evaluation of Pulse Oximetry Accuracy in a Commercial Smartphone and Smartwatch Device During Human Hypoxia Laboratory Testing. Sensors. https://doi.org/10.3390/s25051286

Vancouver

Browne SH, Bernstein M, Bickler PE. Evaluation of Pulse Oximetry Accuracy in a Commercial Smartphone and Smartwatch Device During Human Hypoxia Laboratory Testing. Sensors. 2025. doi:10.3390/s25051286

BibTeX
@article{Browne2025,
  author = {Browne, Sara H and Bernstein, Michael and Bickler, Philip E},
  title = {Evaluation of Pulse Oximetry Accuracy in a Commercial Smartphone and Smartwatch Device During Human Hypoxia Laboratory Testing},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25051286}
}
Kostenlos starten →