EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Smart Force Sensing in Robot Surgery Utilising the Back Electromotive Force

Storm Chabot, Koen Schouten, Bart Van Straten iD, Stefano Pomati, Andres Hunt, Jenny Dankelman iD, Tim Horeman iD

Sensors · 2025

DOI: 10.3390/s25030777 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Storm Chabot und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of BioMechanical Engineering, Delft University of Technology, 2628 CD Delft, The Netherlands. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Robotics. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Netherlands Organisation for Health Research and Development, NWO (Nederlands wetenschappelik Onderzoek).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Since the introduction of robot-assisted laparoscopic surgery, efforts have been made to incorporate force sensing technologies to monitor critical components and to provide force feedback. The advanced laparoscopic robotic system (AdLap RS) is a robotic platform that aims to make robot technology more sustainable through the use of the fully reusable shaft-actuated tip-articulating (SATA) instruments. The SATA instrument driver features electronics and sensors exposed to the sterile environment, which complicate the sterilisation process. The aim of this study was to develop and validate smart sensing in stepper motors using the back electromotive force in a newly developed Smart SATA Driver (SSD), eliminating the need for sensors in the sterile environment.

Methods

The stepper drivers were equipped with TMC2209 ICs featuring StallGuard technology to measure back EMF. The tip was actuated up until a set StallGuard threshold value was reached, at which the resulting tip force was measured. This cycle was repeated ten times for a range of threshold levels. A regression analysis with a power series model was used to determine the quality of the fit.

Results

The SSD is capable of exerting tip forces between 2.4 and 8.2 N. The back EMF force test demonstrated a strong correlation between obtained StallGuard values and measured tip forces. The regression analysis showed an R-squared of 0.95 and a root Mean squared error of 0.4 N.

Discussion

The back EMF force test shows promise for force feedback, but its accuracy limits real-time use due to back EMF fluctuations. Future improvements in motor stability and refining the back EMF model are needed to enable real-time feedback.

Conclusion

The strong correlation during the back EMF force test shows its potential as a low-budget method for detecting motor stalls and estimating tool–tissue forces without the need for sensors in laparoscopic instruments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Robotic Surgical ProceduresRoboticsEquipment DesignLaparoscopyHumans

Stichwörter der Autor:innen

back electromotive forceforce feedbacklaparoscopyrobotic surgerysmart force sensing

Autor:innen und Institutionen

  • Department of BioMechanical Engineering, Delft University of Technology, 2628 CD Delft, The Netherlands.
  • Asensus Surgical, Durham, NC 27703, USA.
  • Department of Precision and Microsystems Engineering, Delft University of Technology, 2628 CD Delft, The Netherlands.

ORCID-Profile: Bart Van Straten, Jenny Dankelman, Tim Horeman

Förderung

  • Netherlands Organisation for Health Research and Development
  • NWO (Nederlands wetenschappelik Onderzoek)

Zitieren

APA

Chabot, S., Schouten, K., Van Straten, B., Pomati, S., Hunt, A., Dankelman, J., & Horeman, T. (2025). Smart Force Sensing in Robot Surgery Utilising the Back Electromotive Force. Sensors. https://doi.org/10.3390/s25030777

Vancouver

Chabot S, Schouten K, Van Straten B, Pomati S, Hunt A, Dankelman J, et al. Smart Force Sensing in Robot Surgery Utilising the Back Electromotive Force. Sensors. 2025. doi:10.3390/s25030777

BibTeX
@article{Chabot2025,
  author = {Chabot, Storm and Schouten, Koen and Van Straten, Bart and Pomati, Stefano and Hunt, Andres and Dankelman, Jenny and Horeman, Tim},
  title = {Smart Force Sensing in Robot Surgery Utilising the Back Electromotive Force},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25030777}
}
Kostenlos starten →