EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

Application of Smart Insoles in Assessing Dynamic Stability in Patients with Chronic Ankle Instability: A Comparative Study

Seonghyun Kang iD, Jaewook Kim iD, Yekwang Kim iD, Juhui Moon iD, Hak Jun Kim iD, Seung-Jong Kim iD

Sensors · 2025

DOI: 10.3390/s25030646 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Sensors, 2025. Verfasst von Seonghyun Kang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Korea University College of Medicine, Seoul 02841, Republic of Korea. Hauptthemen laut MEDLINE: Joint Instability, Ankle Joint, Foot Orthoses. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Health and Welfare, Ministry of Health and Welfare, Republic of Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Chronic ankle instability (CAI), due to its chronic nature and biomechanical complexity, is well-suited for continuous monitoring and tele-rehabilitation using wearable sensor technology. This study assessed whether a smart insole system, equipped with 4 force-sensing resistor sensors and an inertial measurement unit, combined with functional tests and biomechanical indices, could distinguish CAI patients from healthy controls. A total of 21 CAI patients (23.8 ± 5.1 years) and 16 controls (22.62 ± 2.60 years) completed a battery of functional performance tests while wearing the smart insole system. The results showed an increased medial-lateral pressure ratio in the CAI during heel raise (p = 0.031, effect size = 0.82) and hop tests, suggesting an everted foot position. Significant deviations in center-of-pressure trajectory during double-leg heel raises (p = 0.005, effect size = 1.10) suggested asymmetric motion coordination, while compensatory fluctuations of the lifted limb during single-leg balance tests (p = 0.011, effect size = 1.03) were greater in CAI patients. These findings facilitated the development of features to characterize CAI-specific movement patterns. Together, this system shows promise as a quantitative assessment tool for CAI, supporting improved treatment outcomes through tele-rehabilitation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Joint InstabilityAnkle JointFoot OrthosesHumansMaleFemaleAdultYoung AdultBiomechanical PhenomenaWearable Electronic DevicesPostural BalanceChronic DiseaseAnkle

Stichwörter der Autor:innen

ankle functional testsbiomechanicschronic ankle instabilityquantitative assessmentsmart insole

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Korea University College of Medicine, Seoul 02841, Republic of Korea.
  • Department of Orthopaedic Surgery, Korea University Guro Hospital, Seoul 08308, Republic of Korea.

ORCID-Profile: Seonghyun Kang, Jaewook Kim, Yekwang Kim, Juhui Moon, Hak Jun Kim, Seung-Jong Kim

Förderung

  • Ministry of Health and Welfare
  • Ministry of Health and Welfare, Republic of Korea

Zitieren

APA

Kang, S., Kim, J., Kim, Y., Moon, J., Kim, H. J., & Kim, S. J. (2025). Application of Smart Insoles in Assessing Dynamic Stability in Patients with Chronic Ankle Instability: A Comparative Study. Sensors. https://doi.org/10.3390/s25030646

Vancouver

Kang S, Kim J, Kim Y, Moon J, Kim HJ, Kim SJ. Application of Smart Insoles in Assessing Dynamic Stability in Patients with Chronic Ankle Instability: A Comparative Study. Sensors. 2025. doi:10.3390/s25030646

BibTeX
@article{Kang2025,
  author = {Kang, Seonghyun and Kim, Jaewook and Kim, Yekwang and Moon, Juhui and Kim, Hak Jun and Kim, Seung-Jong},
  title = {Application of Smart Insoles in Assessing Dynamic Stability in Patients with Chronic Ankle Instability: A Comparative Study},
  journal = {Sensors},
  year = {2025},
  doi = {10.3390/s25030646}
}
Kostenlos starten →