Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Weixi Zhang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Advanced Processing Technology and Intelligent Manufacturing (Heilongjiang Province), Harbin University of Science and Technology, …. Hauptthemen laut MEDLINE: Breast Neoplasms, Bionics, Breast. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Open Fund for the Key Laboratory of Advanced Processing Technology and Intelligent Manufacturing in Heilongjiang Province, National Natural Science Foundation of China.
Abstract (Originalsprache)
Minimally invasive interventional surgery is commonly used for diagnosing and treating breast cancer, but the high fluidity and deformability of breast tissue reduce intervention accuracy. This study proposes a bionic breast interventional robot that mimics the scorpion’s predation process, actively manipulating tissue deformation to control target displacement and enhance accuracy. The robot’s structure is designed using a modular method, and its kinematics and workspace are analyzed and solved. To address the nonlinear breast tissue deformation problem, a hierarchical tissue method is proposed to simplify the three-dimensional problem into a two-dimensional one. A two-dimensional tissue deformation solver is established based on the minimum energy method for quick resolution. The problem is treated as quasi-static, deriving the displacement relationship between external manipulation points and internal tissue targets. The method of active manipulation of tissue deformation is simulated using MATLAB (2019-b) software to verify the feasibility of the method. Results show maximum errors of 1.7 mm for prostheses and 2.5 mm for in vitro tissues in the X and Y directions. This method improves intervention accuracy in breast surgery and offers a new solution for breast cancer diagnosis and treatment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Key Laboratory of Advanced Processing Technology and Intelligent Manufacturing (Heilongjiang Province), Harbin University of Science and Technology, Harbin 150080, China.
- Postdoctoral Research Center of Instrument Science and Technology, Harbin University of Science and Technology, Harbin 150080, China.
Förderung
- Open Fund for the Key Laboratory of Advanced Processing Technology and Intelligent Manufacturing in Heilongjiang Province
- National Natural Science Foundation of China
Zitieren
APA
Zhang, W., Yu, J., Yu, X., Zhang, Y., & Men, Z. (2024). Study on Bionic Design and Tissue Manipulation of Breast Interventional Robot. Sensors. https://doi.org/10.3390/s24196408
Vancouver
Zhang W, Yu J, Yu X, Zhang Y, Men Z. Study on Bionic Design and Tissue Manipulation of Breast Interventional Robot. Sensors. 2024. doi:10.3390/s24196408
BibTeX
@article{Zhang2024,
author = {Zhang, Weixi and Yu, Jiaxing and Yu, Xiaoyang and Zhang, Yongde and Men, Zhihui},
title = {Study on Bionic Design and Tissue Manipulation of Breast Interventional Robot},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24196408}
}
EUVIMED