Fachartikel in der Zeitschrift Sensors, online erschienen am 26.09.2024. Verfasst von Xinghan Lin und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Artificial Intelligence, Beijing Technology and Business University, Beijing 100048, China. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents the design and validation of a high-precision angular vibration calibration system based on a laser vibrometer, aimed at meeting the high-precision requirements for measuring small angular vibrations. The system primarily consists of a self-driving angular vibration platform and a laser vibrometer. The platform is isolated from ground interference via an air-floating platform and uses a split-type motor to control the platform, generating specific angular vibrations. Detailed simulations of the platform’s modal characteristics and the stability of the spring plates were conducted using the finite element analysis software ANSYS 11. Moreover, fundamental frequency testing and measurement accuracy testing were conducted on the system. Experimental results demonstrate that the system has a fundamental frequency of 2.69 Hz and a maximum measurement error of 0.00172″, confirming the system’s effectiveness in dynamic characteristics, stability, and measurement accuracy. This research provides essential technical support for high-precision angular vibration control in spacecraft.

Schlagwörter

Stichwörter der Autor:innen

accuracy testingangular vibration platform designhigh-precision angular vibration calibration systemsimulation analysis

Autor:innen und Institutionen

  • School of Artificial Intelligence, Beijing Technology and Business University, Beijing 100048, China.

ORCID-Profile: Gang Li

Zitieren

APA

Lin, X., Huang, Z., Guo, K., & Li, G. (2024). The Design and Validation of a High-Precision Angular Vibration Calibration System Based on a Laser Vibrometer. Sensors, 24(19), 6228. https://doi.org/10.3390/s24196228

Vancouver

Lin X, Huang Z, Guo K, Li G. The Design and Validation of a High-Precision Angular Vibration Calibration System Based on a Laser Vibrometer. Sensors. 2024;24(19):6228. doi:10.3390/s24196228

BibTeX
@article{Lin2024,
  author = {Lin, Xinghan and Huang, Zhigang and Guo, Keyou and Li, Gang},
  title = {The Design and Validation of a High-Precision Angular Vibration Calibration System Based on a Laser Vibrometer},
  journal = {Sensors},
  year = {2024},
  volume = {24},
  number = {19},
  pages = {6228},
  doi = {10.3390/s24196228}
}
Kostenlos starten →