A Pressure Sensing Device to Assist in Colonoscopic Procedures to Prevent Perforation—A Case Study
Sensors · 2024
DOI: 10.3390/s24175711 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Se-Eun Kim und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Ajou University, 206 Worldcup-ro, Yeongtong-gu, Suwon-si 16499, Republic of Korea. Hauptthemen laut MEDLINE: Colonoscopy, Pressure, Colon, Intestinal Perforation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Korea Medical Device Development Fund, Korean government (Ministry of Science and ICT, Ministry of Trade, Industry and Energy, Ministry of Health & Welfare, and Ministry of Food and Drug Safety,….
Abstract (Originalsprache)
Colonoscopy has a limited field of view because it relies solely on a small camera attached to the end of the scope and a screen displayed on a monitor. Consequently, the quality and safety of diagnosis and treatment depend on the experience and skills of the gastroenterologist. When a novice attempts to insert the colonoscope during the procedure, excessive pressure can sometimes be applied to the colon wall. This pressure can cause a medical accident known as colonic perforation, which the physician should prevent. We propose an assisting device that senses the pressure applied to the colon wall, analyzes the risk of perforation, and warns the physician in real time. Flexible pressure sensors are attached to the surface of the colonoscope shaft. These sensors measure pressure signals during a colonoscopy procedure. A simple signal processor is used to collect and process the pressure signals. In the experiment, a colonoscope equipped with the proposed device was inserted into a simulated colon made from a colon extracted from a pig. The processed data were visually communicated to the gastroenterologist via displays and light-emitting diodes (LEDs). The device helps the physician continuously monitor and prevent excessive pressure on the colon wall. In this experiment, the device appropriately generated and delivered warnings to help the physicians prevent colonic perforation. In the future, the device is to be improved, and more experiments will be performed in live swine models or humans to confirm its efficacy and safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Ajou University, 206 Worldcup-ro, Yeongtong-gu, Suwon-si 16499, Republic of Korea.
- Department of Internal Medicine, College of Medicine, Soonchunhyang University, 22 Soonchunhyang-ro, Sinchang-myeon, Asan-si 31538, Republic of Korea.
ORCID-Profile: Chang-Ho Jung, Yongho Jeon, Moon Gu Lee
Förderung
- Korea Medical Device Development Fund
- Korean government (Ministry of Science and ICT, Ministry of Trade, Industry and Energy, Ministry of Health & Welfare, and Ministry of Food and Drug Safety,…
Zitieren
APA
Kim, S. E., Kang, Y. J., Jung, C. H., Jeon, Y., Jung, Y., & Lee, M. G. (2024). A Pressure Sensing Device to Assist in Colonoscopic Procedures to Prevent Perforation—A Case Study. Sensors. https://doi.org/10.3390/s24175711
Vancouver
Kim SE, Kang YJ, Jung CH, Jeon Y, Jung Y, Lee MG. A Pressure Sensing Device to Assist in Colonoscopic Procedures to Prevent Perforation—A Case Study. Sensors. 2024. doi:10.3390/s24175711
BibTeX
@article{Kim2024,
author = {Kim, Se-Eun and Kang, Young-Jae and Jung, Chang-Ho and Jeon, Yongho and Jung, Yunho and Lee, Moon Gu},
title = {A Pressure Sensing Device to Assist in Colonoscopic Procedures to Prevent Perforation—A Case Study},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24175711}
}
EUVIMED