Validity and Reliability of the Orthelligent Pro Sensor for Measuring Single-Leg Vertical Jump Height in Healthy Athletic Adults
Sensors · 2024
DOI: 10.3390/s24123699 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Caterina Pasquale und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physiotherapy and Occupational Therapy, University Hospital Zurich, 8091 Zurich, Switzerland. Hauptthemen laut MEDLINE: Leg. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swiss Association of Physiotherapy, Swiss Association of Physiotherpy.
Abstract (Originalsprache)
The Orthelligent Pro sensor is a practicable, portable measuring instrument. This study assessed the validity and reliability of this sensor in measuring single-leg countermovement jumps. Fifty healthy athletic adults participated in two measurement sessions a week apart in time. They performed single-leg countermovement jumps on the force plate while wearing the Orthelligent Pro sensor on their lower leg. During the first measurement session, Tester 1 invited the participants to make three single-leg countermovement jumps; subsequently, Tester 2 did the same. For assessing the sensor’s intratester reliability, Tester 1 again invited the participants to make three single-leg countermovement jumps during the second measurement session. The sensor’s validity was assessed by using the force plate results as the gold standard. To determinate the agreement between two measurements, Bland–Altman plots were created. The intertester reliability (ICC = 0.99; 0.97) and intratester reliability (ICC = 0.96; 0.82) were both excellent. The validity calculated (i) on the basis of the mean value of three jumps and (ii) on the basis of the maximum value of three jumps was very high, but it showed a systematic error. Taking this error into account, physiotherapists can use the Orthelligent Pro sensor as a valid and reliable instrument for measuring the jump height of countermovement jumps.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Physiotherapy and Occupational Therapy, University Hospital Zurich, 8091 Zurich, Switzerland.
- Directorate of Research and Education, Physiotherapy Occupational Therapy Research Center, University Hospital Zurich, 8091 Zurich, Switzerland.
- Faculty of Medicine, University of Zurich, 8006 Zurich, Switzerland.
- Department of Pulmonology, University Hospital Zurich, 8091 Zurich, Switzerland.
- Department of Sport and Exercise Science, University of Salzburg, 5020 Salzburg, Austria.
- Red Bull Athlete Performance Center, 5303 Salzburg, Austria.
ORCID-Profile: Pierrette Baschung Pfister, Manuel Kuhn
Förderung
- Swiss Association of Physiotherapy
- Swiss Association of Physiotherpy
Zitieren
APA
Pasquale, C., Baschung Pfister, P., Kuhn, M., & Stöggl, T. (2024). Validity and Reliability of the Orthelligent Pro Sensor for Measuring Single-Leg Vertical Jump Height in Healthy Athletic Adults. Sensors. https://doi.org/10.3390/s24123699
Vancouver
Pasquale C, Baschung Pfister P, Kuhn M, Stöggl T. Validity and Reliability of the Orthelligent Pro Sensor for Measuring Single-Leg Vertical Jump Height in Healthy Athletic Adults. Sensors. 2024. doi:10.3390/s24123699
BibTeX
@article{Pasquale2024,
author = {Pasquale, Caterina and Baschung Pfister, Pierrette and Kuhn, Manuel and Stöggl, Thomas},
title = {Validity and Reliability of the Orthelligent Pro Sensor for Measuring Single-Leg Vertical Jump Height in Healthy Athletic Adults},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24123699}
}
EUVIMED