Optimization of Surface Functionalizations for Ring Resonator-Based Biosensors
Sensors · 14.05.2024 · 24 (10) · 3107
DOI: 10.3390/s24103107 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 14.05.2024. Verfasst von Niccolò Ardoino und 8 weiteren Autor:innen. Beteiligte Einrichtung: FTH S.r.l., Via Sommarive 18, I-38123 Trento, Italy. Hauptthemen laut MEDLINE: Biosensing Techniques, Aptamers, Nucleotide, Silanes, Thrombin. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Autonomous Province of Trento (Italy).
Abstract (Originalsprache)
Liquid biopsy is expected to become widespread in the coming years thanks to point of care devices, which can include label-free biosensors. The surface functionalization of biosensors is a crucial aspect that influences their overall performance, resulting in the accurate, sensitive, and specific detection of target molecules. Here, the surface of a microring resonator (MRR)-based biosensor was functionalized for the detection of protein biomarkers. Among the several existing functionalization methods, a strategy based on aptamers and mercaptosilanes was selected as the most highly performing approach. All steps of the functionalization protocol were carefully characterized and optimized to obtain a suitable protocol to be transferred to the final biosensor. The functionalization protocol comprised a preliminary plasma treatment aimed at cleaning and activating the surface for the subsequent silanization step. Different plasma treatments as well as different silanes were tested in order to covalently bind aptamers specific to different biomarker targets, i.e., C-reactive protein, SARS-CoV-2 spike protein, and thrombin. Argon plasma and 1% v/v mercaptosilane were found as the most suitable for obtaining a homogeneous layer apt to aptamer conjugation. The aptamer concentration and time for immobilization were optimized, resulting in 1 µM and 3 h, respectively. A final passivation step based on mercaptohexanol was also implemented. The functionalization protocol was then evaluated for the detection of thrombin with a photonic biosensor based on microring resonators. The preliminary results identified the successful recognition of the correct target as well as some limitations of the developed protocol in real measurement conditions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- FTH S.r.l., Via Sommarive 18, I-38123 Trento, Italy.
- Center for Sensors & Devices, Fondazione Bruno Kessler, Via Sommarive 18, I-38123 Trento, Italy.
- Istituto di Biofisica, Consiglio Nazionale delle Ricerche, Via alla Cascata 56/C, I-38123 Trento, Italy.
- Department of Physics, University of Trento, Via Sommarive 14, Povo, I-38123 Trento, Italy.
- Indivenire S.r.l., Via Sommarive 18, I-38123 Trento, Italy.
- Department of Engineering, University of Campania "Luigi Vanvitelli", Via Roma 29, I-81031 Aversa, Italy.
ORCID-Profile: Niccolò Ardoino, Lorenzo Lunelli, Georg Pucker, Lia Vanzetti, Rachele Favaretto, Laura Pasquardini, Cecilia Pederzolli, Carlo Guardiani, Cristina Potrich
Förderung
- Autonomous Province of Trento (Italy)
Zitieren
APA
Ardoino, N., Lunelli, L., Pucker, G., Vanzetti, L., Favaretto, R., Pasquardini, L., Pederzolli, C., Guardiani, C., & Potrich, C. (2024). Optimization of Surface Functionalizations for Ring Resonator-Based Biosensors. Sensors, 24(10), 3107. https://doi.org/10.3390/s24103107
Vancouver
Ardoino N, Lunelli L, Pucker G, Vanzetti L, Favaretto R, Pasquardini L, et al. Optimization of Surface Functionalizations for Ring Resonator-Based Biosensors. Sensors. 2024;24(10):3107. doi:10.3390/s24103107
BibTeX
@article{Ardoino2024,
author = {Ardoino, Niccolò and Lunelli, Lorenzo and Pucker, Georg and Vanzetti, Lia and Favaretto, Rachele and Pasquardini, Laura and Pederzolli, Cecilia and Guardiani, Carlo and Potrich, Cristina},
title = {Optimization of Surface Functionalizations for Ring Resonator-Based Biosensors},
journal = {Sensors},
year = {2024},
volume = {24},
number = {10},
pages = {3107},
doi = {10.3390/s24103107}
}
EUVIMED