Evaluation of Hand Action Classification Performance Using Machine Learning Based on Signals from Two sEMG Electrodes

Hope O Shaw iD, Kirstie M Devin iD, Jinghua Tang iD, Liudi Jiang

Sensors · 2024

DOI: 10.3390/s24082383 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Hope O Shaw und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Engineering, Faculty of Engineering and Physical Sciences, University of Southampton, Southampton SO17 1BJ, UK. Hauptthemen laut MEDLINE: Electromyography, Hand, Machine Learning, Electrodes. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ESPRC, UK Engineering and Physical Sciences Research Council (EPSRC).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Classification-based myoelectric control has attracted significant interest in recent years, leading to prosthetic hands with advanced functionality, such as multi-grip hands. Thus far, high classification accuracies have been achieved by increasing the number of surface electromyography (sEMG) electrodes or adding other sensing mechanisms. While many prescribed myoelectric hands still adopt two-electrode sEMG systems, detailed studies on signal processing and classification performance are still lacking. In this study, nine able-bodied participants were recruited to perform six typical hand actions, from which sEMG signals from two electrodes were acquired using a Delsys Trigno Research+ acquisition system. Signal processing and machine learning algorithms, specifically, linear discriminant analysis (LDA), k-nearest neighbors (KNN), and support vector machines (SVM), were used to study classification accuracies. Overall classification accuracy of 93 ± 2%, action-specific accuracy of 97 ± 2%, and F1-score of 87 ± 7% were achieved, which are comparable with those reported from multi-electrode systems. The highest accuracies were achieved using SVM algorithm compared to LDA and KNN algorithms. A logarithmic relationship between classification accuracy and number of features was revealed, which plateaued at five features. These comprehensive findings may potentially contribute to signal processing and machine learning strategies for commonly prescribed myoelectric hand systems with two sEMG electrodes to further improve functionality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectromyographyHandMachine LearningElectrodesSignal Processing, Computer-AssistedSupport Vector MachineAlgorithmsHumansMaleAdultFemaleDiscriminant AnalysisYoung Adult

Stichwörter der Autor:innen

classificationhand actionsmachine learningmyoelectric prostheticssEMGupper limb

Autor:innen und Institutionen

  • School of Engineering, Faculty of Engineering and Physical Sciences, University of Southampton, Southampton SO17 1BJ, UK.

ORCID-Profile: Hope O Shaw, Kirstie M Devin, Jinghua Tang

Förderung

  • ESPRC
  • UK Engineering and Physical Sciences Research Council (EPSRC)

Zitieren

APA

Shaw, H. O., Devin, K. M., Tang, J., & Jiang, L. (2024). Evaluation of Hand Action Classification Performance Using Machine Learning Based on Signals from Two sEMG Electrodes. Sensors. https://doi.org/10.3390/s24082383

Vancouver

Shaw HO, Devin KM, Tang J, Jiang L. Evaluation of Hand Action Classification Performance Using Machine Learning Based on Signals from Two sEMG Electrodes. Sensors. 2024. doi:10.3390/s24082383

BibTeX
@article{Shaw2024,
  author = {Shaw, Hope O and Devin, Kirstie M and Tang, Jinghua and Jiang, Liudi},
  title = {Evaluation of Hand Action Classification Performance Using Machine Learning Based on Signals from Two sEMG Electrodes},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24082383}
}
Kostenlos starten →