Vertical Center-of-Mass Braking and Motor Performance during Gait Initiation in Young Healthy Adults, Elderly Healthy Adults, and Patients with Parkinson’s Disease: A Comparison of Force-Plate and Markerless Motion Capture Systems
Sensors · 2024
DOI: 10.3390/s24041302 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Arnaud Simonet und 3 weiteren Autor:innen. Beteiligte Einrichtung: LADAPT Loiret, Centre de Soins de Suite et de Réadaptation, 45200 Amilly, France. Hauptthemen laut MEDLINE: Parkinson Disease. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Association Nationale de la Recherche et de la Technologie, LADAPT Loiret.
Abstract (Originalsprache)
Background
This study tested the agreement between a markerless motion capture system and force-plate system (“gold standard”) to quantify stability control and motor performance during gait initiation.
Methods
Healthy adults (young and elderly) and patients with Parkinson’s disease performed gait initiation series at spontaneous and maximal velocity on a system of two force-plates placed in series while being filmed by a markerless motion capture system. Signals from both systems were used to compute the peak of forward center-of-mass velocity (indicator of motor performance) and the braking index (indicator of stability control).
Results
Descriptive statistics indicated that both systems detected between-group differences and velocity effects similarly, while a Bland–Altman plot analysis showed that mean biases of both biomechanical indicators were virtually zero in all groups and conditions. Bayes factor 01 indicated strong (braking index) and moderate (motor performance) evidence that both systems provided equivalent values. However, a trial-by-trial analysis of Bland–Altman plots revealed the possibility of differences >10% between the two systems.
Conclusion
Although non-negligible differences do occur, a markerless motion capture system appears to be as efficient as a force-plate system in detecting Parkinson’s disease and velocity condition effects on the braking index and motor performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- LADAPT Loiret, Centre de Soins de Suite et de Réadaptation, 45200 Amilly, France.
- CIAMS, Université Paris-Saclay, 91190 Paris, France.
- CIAMS, Université d'Orléans, 45067 Orléans, France.
- Département de Chirurgie Orthopédique, Université Libre de Bruxelles, 1050 Bruxelles, Belgium.
ORCID-Profile: Arnaud Delafontaine, Eric Yiou
Förderung
- Association Nationale de la Recherche et de la Technologie
- LADAPT Loiret
Zitieren
APA
Simonet, A., Delafontaine, A., Fourcade, P., & Yiou, E. (2024). Vertical Center-of-Mass Braking and Motor Performance during Gait Initiation in Young Healthy Adults, Elderly Healthy Adults, and Patients with Parkinson’s Disease: A Comparison of Force-Plate and Markerless Motion Capture Systems. Sensors. https://doi.org/10.3390/s24041302
Vancouver
Simonet A, Delafontaine A, Fourcade P, Yiou E. Vertical Center-of-Mass Braking and Motor Performance during Gait Initiation in Young Healthy Adults, Elderly Healthy Adults, and Patients with Parkinson’s Disease: A Comparison of Force-Plate and Markerless Motion Capture Systems. Sensors. 2024. doi:10.3390/s24041302
BibTeX
@article{Simonet2024,
author = {Simonet, Arnaud and Delafontaine, Arnaud and Fourcade, Paul and Yiou, Eric},
title = {Vertical Center-of-Mass Braking and Motor Performance during Gait Initiation in Young Healthy Adults, Elderly Healthy Adults, and Patients with Parkinson’s Disease: A Comparison of Force-Plate and Markerless Motion Capture Systems},
journal = {Sensors},
year = {2024},
doi = {10.3390/s24041302}
}
EUVIMED