Fachartikel in der Zeitschrift Sensors, online erschienen am 07.02.2024. Verfasst von Yueyu Lin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Life Science Trace Detection Laboratory, Department of Analytical Chemistry and Chemometrics, Institute for Molecules and Materials, Radboud Universi…. Hauptthemen laut MEDLINE: Carbon Dioxide, Methane. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dutch Research Council, the Netherlands Organisation for Scientific Research (NWO) “Synergia-SYstem change for New Ecology-based and Resource efficient Growth with high tech In Agricu….

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The presence of an elevated amount of methane (CH4) in exhaled breath can be used as a non-invasive tool to monitor certain health conditions. A compact, inexpensive and transportable CH4 sensor is thus very interesting for this purpose. In addition, if the sensor is also able to simultaneously measure carbon dioxide (CO2), one can extract the end-tidal concentration of exhaled CH4. Here, we report on such a sensor based on a commercial detection module using tunable diode laser absorption spectroscopy. It was found that the measured CH4/CO2 values exhibit a strong interference with water vapor. Therefore, correction functions were experimentally identified and validated for both CO2 and CH4. A custom-built breath sampler was developed and tested with the sensor for real-time measurements of CH4 and CO2 in exhaled breath. As a result, the breath sensor demonstrated the capability of accurately measuring the exhaled CH4 and CO2 profiles in real-time. We obtained minimum detection limits of ~80 ppbv for CH4 and ~700 ppmv for CO2 in 1.5 s measurement time.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Carbon DioxideMethaneHumansSpectrum Analysis

Stichwörter der Autor:innen

breathcarbon dioxidecorrection functionsmethanereal-time measurements

Substanzen

Carbon DioxideMethane

Autor:innen und Institutionen

  • Life Science Trace Detection Laboratory, Department of Analytical Chemistry and Chemometrics, Institute for Molecules and Materials, Radboud University, 6525 AJ Nijmegen, The Netherlands.

ORCID-Profile: Yueyu Lin, Dexter Manalili, Simona M Cristescu

Förderung

  • Dutch Research Council
  • the Netherlands Organisation for Scientific Research (NWO) “Synergia-SYstem change for New Ecology-based and Resource efficient Growth with high tech In Agricu…

Zitieren

APA

Lin, Y., Manalili, D., Khodabakhsh, A., & Cristescu, S. M. (2024). Real-Time Measurement of CH4 in Human Breath Using a Compact CH4/CO2 Sensor. Sensors, 24(4), 1077. https://doi.org/10.3390/s24041077

Vancouver

Lin Y, Manalili D, Khodabakhsh A, Cristescu SM. Real-Time Measurement of CH4 in Human Breath Using a Compact CH4/CO2 Sensor. Sensors. 2024;24(4):1077. doi:10.3390/s24041077

BibTeX
@article{Lin2024,
  author = {Lin, Yueyu and Manalili, Dexter and Khodabakhsh, Amir and Cristescu, Simona M},
  title = {Real-Time Measurement of CH4 in Human Breath Using a Compact CH4/CO2 Sensor},
  journal = {Sensors},
  year = {2024},
  volume = {24},
  number = {4},
  pages = {1077},
  doi = {10.3390/s24041077}
}
Kostenlos starten →