EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Analyzing the Automatic Power Level Control Effect of a Signal Generator in RF Power Sensor Calibration by a Direct Comparison Transfer Method and a Millimeter Wave Application

Erkan Danaci iD

Sensors · 18.01.2024 · 24 (2) · 609

DOI: 10.3390/s24020609 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 18.01.2024. Verfasst von Erkan Danaci. Beteiligte Einrichtung: TUBITAK National Metrology Institute (UME), P.O. Box 54, Kocaeli 41470, Türkiye. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: TÜBİTAK UME.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Most calibration laboratories prefer the Direct Comparison Transfer Method (DCTM) for a reliable and accurate calibration of power sensors in the radio frequency (RF) scope. Most studies suggest using this calibration method, with its automatic power level control (APLC) of RF signal generators. The APLC is preferred to keep the output power level of the signal generator the same, while the power sensor is calibrated and the reference power sensor is connected to the measurement system. The known APLC mechanisms are also explained for the DCTM, and a comparison of the calibration factor values carried out with and without the automatic power level control process in the DCTM is also given in this study. RF power sensor calibrations with coaxial and waveguide connector types are examined with DCTM in this study as well. This study shows that the DCTM, unless with APLC, should be applied for the waveguide power sensor’s calibration at millimeter wave frequencies.

Schlagwörter

Stichwörter der Autor:innen

automatic level control of RF sourcecoaxial RF power sensor calibrationdirect comparison transfer methodmillimeter wave power sensor calibration

Autor:innen und Institutionen

  • TUBITAK National Metrology Institute (UME), P.O. Box 54, Kocaeli 41470, Türkiye.

ORCID-Profile: Erkan Danaci

Förderung

  • TÜBİTAK UME

Zitieren

APA

Danaci, E. (2024). Analyzing the Automatic Power Level Control Effect of a Signal Generator in RF Power Sensor Calibration by a Direct Comparison Transfer Method and a Millimeter Wave Application. Sensors, 24(2), 609. https://doi.org/10.3390/s24020609

Vancouver

Danaci E. Analyzing the Automatic Power Level Control Effect of a Signal Generator in RF Power Sensor Calibration by a Direct Comparison Transfer Method and a Millimeter Wave Application. Sensors. 2024;24(2):609. doi:10.3390/s24020609

BibTeX
@article{Danaci2024,
  author = {Danaci, Erkan},
  title = {Analyzing the Automatic Power Level Control Effect of a Signal Generator in RF Power Sensor Calibration by a Direct Comparison Transfer Method and a Millimeter Wave Application},
  journal = {Sensors},
  year = {2024},
  volume = {24},
  number = {2},
  pages = {609},
  doi = {10.3390/s24020609}
}
Kostenlos starten →