Fachartikel in der Zeitschrift Sensors, 2024. Verfasst von Na Jin Seo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Rehabilitation Sciences, College of Health Professions, Medical University of South Carolina, Charleston, SC 29425, USA. Hauptthemen laut MEDLINE: Stroke, Wearable Electronic Devices. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIGMS NIH HHS, NICHD NIH HHS, NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Many stroke survivors experience persistent upper extremity impairment that limits performance in activities of daily living. Upper limb recovery requires high repetitions of task-specific practice. Stroke survivors are often prescribed task practices at home to supplement rehabilitation therapy. A poor quality of task practices, such as the use of compensatory movement patterns, results in maladaptive neuroplasticity and suboptimal motor recovery. There currently lacks a tool for the remote monitoring of movement quality of stroke survivors’ task practices at home. The objective of this study was to evaluate the feasibility of classifying movement quality at home using a wearable IMU. Nineteen stroke survivors wore an IMU sensor on the paretic wrist and performed four functional upper limb tasks in the lab and later at home while videorecording themselves. The lab data served as reference data to classify home movement quality using dynamic time warping. Incorrect and correct movement quality was labeled by a therapist. The home task practice movement quality was classified with an accuracy of 92% and F1 score of 0.95 for all tasks combined. Movement types contributing to misclassification were further investigated. The results support the feasibility of a home movement quality monitoring system to assist with upper limb rehabilitation post stroke.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

StrokeWearable Electronic DevicesHumansActivities of Daily LivingUpper ExtremityWristSurvivors

Stichwörter der Autor:innen

accelerometerhome monitoringinertial measurement unit (IMU)machine learningrehabilitationremote monitoringstrokeupper extremitywearable sensor

Autor:innen und Institutionen

  • Department of Rehabilitation Sciences, College of Health Professions, Medical University of South Carolina, Charleston, SC 29425, USA.
  • Department of Health Sciences and Research, College of Health Professions, Medical University of South Carolina, Charleston, SC 29425, USA.
  • Ralph H. Johnson VA Health Care System, Charleston, SC 29401, USA.

ORCID-Profile: Na Jin Seo, Gabrielle Scronce

Förderung

  • NIGMS NIH HHS
  • NICHD NIH HHS
  • NIH HHS
  • NIH/NICHD P2C HD101913 Pilot
  • NIH/NIGMS P20 GM109040

Zitieren

APA

Seo, N. J., Coupland, K., Finetto, C., & Scronce, G. (2024). Wearable Sensor to Monitor Quality of Upper Limb Task Practice for Stroke Survivors at Home. Sensors. https://doi.org/10.3390/s24020554

Vancouver

Seo NJ, Coupland K, Finetto C, Scronce G. Wearable Sensor to Monitor Quality of Upper Limb Task Practice for Stroke Survivors at Home. Sensors. 2024. doi:10.3390/s24020554

BibTeX
@article{Seo2024,
  author = {Seo, Na Jin and Coupland, Kristen and Finetto, Christian and Scronce, Gabrielle},
  title = {Wearable Sensor to Monitor Quality of Upper Limb Task Practice for Stroke Survivors at Home},
  journal = {Sensors},
  year = {2024},
  doi = {10.3390/s24020554}
}
Kostenlos starten →