Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring: Insights from Pig and Human Cadaver Studies

Salar Tayebi iD, Tim McKinney, Cynthia McKinney, Dipak Delvadia, Marc-Alan Levine, Edward S Spofford, Luca Malbrain, Johan Stiens iD, Wojciech Dabrowski, Manu L N G Malbrain iD

Sensors · 2023

DOI: 10.3390/s23218806 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Salar Tayebi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics and Informatics, Vrije Universiteit Brussel, 1050 Brussels, Belgium. Hauptthemen laut MEDLINE: Intra-Abdominal Hypertension. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sentinel Medical Technologies.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Intra-abdominal pressure (IAP) monitoring is crucial for the detection and prevention of intra-abdominal hypertension (IAH) and abdominal compartment syndrome (ACS). In the 1970s, air-filled catheters (AFCs) for urodynamic studies were introduced as a solution to overcome the limitations of water-perfused catheters. Recent studies have shown that for correct IAP measurement with traditional AFC, the bladder needs to be primed with 25 mL of saline solution to allow pressure wave transmission to the transducer outside of the body, which limits continuous IAP monitoring.

Methods

In this study, a novel triple balloon, air-filled TraumaGuard (TG) catheter system from Sentinel Medical Technologies (Jacksonville, FL, USA) with a unique balloon-in-balloon design was evaluated in a porcine and cadaver model of IAH via laparoscopy (IAPgold).

Results

In total, 27 and 86 paired IAP measurements were performed in two pigs and one human cadaver, respectively. The mean IAPTG was 20.7 ± 10.7 mmHg compared to IAPgold of 20.3 ± 10.3 mmHg in the porcine study. In the cadaver investigation, the mean IAPTG was 15.6 ± 10.8 mmHg compared to IAPgold of 14.4 ± 10.4 mmHg. The correlation, concordance, bias, precision, limits of agreement, and percentage error were all in accordance with the WSACS (Abdominal Compartment Society) recommendations and guidelines for research.

Conclusions

These findings support the use of the TG catheter for continuous IAP monitoring, providing early detection of elevated IAP, thus enabling the potential for prevention of IAH and ACS. Confirmation studies with the TraumaGuard system in critically ill patients are warranted to further validate these findings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intra-Abdominal HypertensionHumansAnimalsSwineCritical IllnessCatheters

Stichwörter der Autor:innen

abdominal compartment syndromeair-filled catheterballoon-in-balloon designcontinuous measurementintra-abdominal hypertensionintra-abdominal pressuresupine position

Autor:innen und Institutionen

  • Department of Electronics and Informatics, Vrije Universiteit Brussel, 1050 Brussels, Belgium.
  • College of Medicine, Drexel University, Philadelphia, PA 19129, USA.
  • SGU School of Medicine, Danbury University Hospital, Danbury, CT 06810, USA.
  • Cricket Innovations, Pottstown, PA 19465, USA.
  • Faculty of Medicine, Katholieke Universiteit Leuven, 3000 Leuven, Belgium.
  • First Department of Anaesthesiology and Intensive Therapy, Medical University of Lublin, 20-954 Lublin, Poland.

ORCID-Profile: Salar Tayebi, Johan Stiens, Manu L N G Malbrain

Förderung

  • Sentinel Medical Technologies

Zitieren

APA

Tayebi, S., McKinney, T., McKinney, C., Delvadia, D., Levine, M. A., Spofford, E. S., Malbrain, L., Stiens, J., Dabrowski, W., & Malbrain, M. L. N. G. (2023). Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring: Insights from Pig and Human Cadaver Studies. Sensors. https://doi.org/10.3390/s23218806

Vancouver

Tayebi S, McKinney T, McKinney C, Delvadia D, Levine MA, Spofford ES, et al. Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring: Insights from Pig and Human Cadaver Studies. Sensors. 2023. doi:10.3390/s23218806

BibTeX
@article{Tayebi2023,
  author = {Tayebi, Salar and McKinney, Tim and McKinney, Cynthia and Delvadia, Dipak and Levine, Marc-Alan and Spofford, Edward S and Malbrain, Luca and Stiens, Johan and Dabrowski, Wojciech and Malbrain, Manu L N G},
  title = {Evaluation of the TraumaGuard Balloon-in-Balloon Catheter Design for Intra-Abdominal Pressure Monitoring: Insights from Pig and Human Cadaver Studies},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23218806}
}
Kostenlos starten →