EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Optimal Frequency and Wireless Power Budget for Miniature Receivers in Obese People

Tom Van de Steene iD, Emmeric Tanghe, Luc Martens, Carmine Garripoli, Stefano Stanzione, Wout Joseph iD

Sensors · 2023

DOI: 10.3390/s23198084 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Tom Van de Steene und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Information Technology, Ghent University/imec, B-9052 Ghent, Belgium. Hauptthemen laut MEDLINE: Prostheses and Implants, Wireless Technology. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 66 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Foundation - Flanders, Imec AAA-ingestibles project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study investigates wireless power transfer for deep in-body receivers, determining the optimal frequency, power budget, and design for the transmitter and receiver. In particular, the focus is on small, in-body receivers at large depths up to 20 cm for obese patients. This enables long-term monitoring of the gastrointestinal tract for all body types. Numerical simulations are used to investigate power transfer and losses as a function of frequency and to find the optimal design at the selected frequency for an obese body model. From all ISM-frequencies in the investigated range (1 kHz–10 GHz), the value of 13.56 MHz yields the best performance. This optimum corresponds to the transition from dominant copper losses in conductors to dominant losses in conductive tissue. At this frequency, a transmitting and receiving coil are designed consisting of 12 and 23 windings, respectively. With a power transfer efficiency of 2.70×10−5, 18 µW can be received for an input power of 0.68 W while still satisfying exposure guidelines. The power transfer is validated by measurements. For the first time, efficiency values and the power budget are reported for WPT through 20 cm of tissue to mm sized receivers. Compared to WPT at higher frequencies, as commonly used for small receivers, the proposed system is more suitable for WPT to large depths in-body and comes with the advantage that no focusing is required, which can accommodate multiple receivers and uncertainty about receiver location more easily. The received power allows long-term sensing in the gastrointestinal tract by, e.g., temperature, pressure, and pH sensors, motility sensing, or even gastric stimulation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostheses and ImplantsWireless TechnologyHumansElectric Power SuppliesEquipment DesignElectric Conductivity

Stichwörter der Autor:innen

coil designdeep implantsingestiblesminiature implantable medical devicesnumerical simulationsresonant inductive power transferwireless power transfer

Autor:innen und Institutionen

  • Department of Information Technology, Ghent University/imec, B-9052 Ghent, Belgium.
  • Holst Centre/imec, 5656 AE Eindhoven, The Netherlands.

ORCID-Profile: Tom Van de Steene, Wout Joseph

Förderung

  • Research Foundation - Flanders
  • Imec AAA-ingestibles project

Zitieren

APA

Van de Steene, T., Tanghe, E., Martens, L., Garripoli, C., Stanzione, S., & Joseph, W. (2023). Optimal Frequency and Wireless Power Budget for Miniature Receivers in Obese People. Sensors. https://doi.org/10.3390/s23198084

Vancouver

Van de Steene T, Tanghe E, Martens L, Garripoli C, Stanzione S, Joseph W. Optimal Frequency and Wireless Power Budget for Miniature Receivers in Obese People. Sensors. 2023. doi:10.3390/s23198084

BibTeX
@article{VandeSteene2023,
  author = {Van de Steene, Tom and Tanghe, Emmeric and Martens, Luc and Garripoli, Carmine and Stanzione, Stefano and Joseph, Wout},
  title = {Optimal Frequency and Wireless Power Budget for Miniature Receivers in Obese People},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23198084}
}
Kostenlos starten →