EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Best Practices for Body Temperature Measurement with Infrared Thermography: External Factors Affecting Accuracy

Siavash Mazdeyasna iD, Pejman Ghassemi iD, Quanzeng Wang iD

Sensors · 21.09.2023 · 23 (18) · 8011

DOI: 10.3390/s23188011 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 21.09.2023. Verfasst von Siavash Mazdeyasna und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD 20993, USA. Hauptthemen laut MEDLINE: Thermography, Body Temperature. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 93 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the U.S. Food and Drug Administration’s Emergency Preparedness program.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Infrared thermographs (IRTs) are commonly used during disease pandemics to screen individuals with elevated body temperature (EBT). To address the limited research on external factors affecting IRT accuracy, we conducted benchtop measurements and computer simulations with two IRTs, with or without an external temperature reference source (ETRS) for temperature compensation. The combination of an IRT and an ETRS forms a screening thermograph (ST). We investigated the effects of viewing angle (θ, 0–75°), ETRS set temperature (TETRS, 30–40 °C), ambient temperature (Tatm, 18–32 °C), relative humidity (RH, 15–80%), and working distance (d, 0.4–2.8 m). We discovered that STs exhibited higher accuracy compared to IRTs alone. Across the tested ranges of Tatm and RH, both IRTs exhibited absolute measurement errors of less than 0.97 °C, while both STs maintained absolute measurement errors of less than 0.12 °C. The optimal TETRS for EBT detection was 36–37 °C. When θ was below 30°, the two STs underestimated calibration source (CS) temperature (TCS) of less than 0.05 °C. The computer simulations showed absolute temperature differences of up to 0.28 °C and 0.04 °C between estimated and theoretical temperatures for IRTs and STs, respectively, considering d of 0.2–3.0 m, Tatm of 15–35 °C, and RH of 5–95%. The results highlight the importance of precise calibration and environmental control for reliable temperature readings and suggest proper ranges for these factors, aiming to enhance current standard documents and best practice guidelines. These insights enhance our understanding of IRT performance and their sensitivity to various factors, thereby facilitating the development of best practices for accurate EBT measurement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ThermographyBody TemperatureHumansInfrared RaysComputer SimulationHumidityCalibration

Stichwörter der Autor:innen

ISO/TR 13154accuracyambient temperatureatmosphere transmittanceelevated body temperatureenvironmental effectsexternal temperature reference sourceinfrared thermographrelative humiditythermographyviewing angle

Autor:innen und Institutionen

  • Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD 20993, USA.

ORCID-Profile: Siavash Mazdeyasna, Pejman Ghassemi, Quanzeng Wang

Förderung

  • the U.S. Food and Drug Administration’s Emergency Preparedness program

Zitieren

APA

Mazdeyasna, S., Ghassemi, P., & Wang, Q. (2023). Best Practices for Body Temperature Measurement with Infrared Thermography: External Factors Affecting Accuracy. Sensors, 23(18), 8011. https://doi.org/10.3390/s23188011

Vancouver

Mazdeyasna S, Ghassemi P, Wang Q. Best Practices for Body Temperature Measurement with Infrared Thermography: External Factors Affecting Accuracy. Sensors. 2023;23(18):8011. doi:10.3390/s23188011

BibTeX
@article{Mazdeyasna2023,
  author = {Mazdeyasna, Siavash and Ghassemi, Pejman and Wang, Quanzeng},
  title = {Best Practices for Body Temperature Measurement with Infrared Thermography: External Factors Affecting Accuracy},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {18},
  pages = {8011},
  doi = {10.3390/s23188011}
}
Kostenlos starten →