Sensor Selection for Tidal Volume Determination via Linear Regression—Impact of Lasso versus Ridge Regression
Sensors · 25.08.2023 · 23 (17) · 7407
DOI: 10.3390/s23177407 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, online erschienen am 25.08.2023. Verfasst von Bernhard Laufer und 8 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Technical Medicine (ITeM), Furtwangen University, 78054 Villingen-Schwenningen, Germany. Hauptthemen laut MEDLINE: Benchmarking, Home Care Services. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: German Federal Ministry for Economic Affairs and Climate Action, German Federal Ministry of Education and Research, European Commission.
Abstract (Originalsprache)
The measurement of respiratory volume based on upper body movements by means of a smart shirt is increasingly requested in medical applications. This research used upper body surface motions obtained by a motion capture system, and two regression methods to determine the optimal selection and placement of sensors on a smart shirt to recover respiratory parameters from benchmark spirometry values. The results of the two regression methods (Ridge regression and the least absolute shrinkage and selection operator (Lasso)) were compared. This work shows that the Lasso method offers advantages compared to the Ridge regression, as it provides sparse solutions and is more robust to outliers. However, both methods can be used in this application since they lead to a similar sensor subset with lower computational demand (from exponential effort for full exhaustive search down to the order of O (n2)). A smart shirt for respiratory volume estimation could replace spirometry in some cases and would allow for a more convenient measurement of respiratory parameters in home care or hospital settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Technical Medicine (ITeM), Furtwangen University, 78054 Villingen-Schwenningen, Germany.
- Department of Mechanical Engineering, University of Canterbury, Christchurch 8041, New Zealand.
- School of Mathematics and Statistics, University of Canterbury, Christchurch 8041, New Zealand.
- Innovation Center Computer Assisted Surgery (ICCAS), University of Leipzig, 04109 Leipzig, Germany.
- Department of Microsystems Engineering, University of Freiburg, 79085 Freiburg, Germany.
ORCID-Profile: Bernhard Laufer, Paul D Docherty, Rua Murray, Nour Aldeen Jalal, Fabian Hoeflinger, Leonhard Reindl, Knut Moeller
Förderung
- German Federal Ministry for Economic Affairs and Climate Action
- German Federal Ministry of Education and Research
- European Commission
Zitieren
APA
Laufer, B., Docherty, P. D., Murray, R., Krueger-Ziolek, S., Jalal, N. A., Hoeflinger, F., Rupitsch, S. J., Reindl, L., & Moeller, K. (2023). Sensor Selection for Tidal Volume Determination via Linear Regression—Impact of Lasso versus Ridge Regression. Sensors, 23(17), 7407. https://doi.org/10.3390/s23177407
Vancouver
Laufer B, Docherty PD, Murray R, Krueger-Ziolek S, Jalal NA, Hoeflinger F, et al. Sensor Selection for Tidal Volume Determination via Linear Regression—Impact of Lasso versus Ridge Regression. Sensors. 2023;23(17):7407. doi:10.3390/s23177407
BibTeX
@article{Laufer2023,
author = {Laufer, Bernhard and Docherty, Paul D and Murray, Rua and Krueger-Ziolek, Sabine and Jalal, Nour Aldeen and Hoeflinger, Fabian and Rupitsch, Stefan J and Reindl, Leonhard and Moeller, Knut},
title = {Sensor Selection for Tidal Volume Determination via Linear Regression—Impact of Lasso versus Ridge Regression},
journal = {Sensors},
year = {2023},
volume = {23},
number = {17},
pages = {7407},
doi = {10.3390/s23177407}
}
EUVIMED