Evaluation of Upper Body and Lower Limbs Kinematics through an IMU-Based Medical System: A Comparative Study with the Optoelectronic System

Serena Cerfoglio iD, Paolo Capodaglio iD, Paolo Rossi iD, Ilaria Conforti iD, Valentina D'Angeli, Elia Milani, Manuela Galli iD, Veronica Cimolin iD

Sensors · 05.07.2023 · 23 (13) · 6156

DOI: 10.3390/s23136156 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 05.07.2023. Verfasst von Serena Cerfoglio und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics, Information and Bioengineering, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milano, Italy. Hauptthemen laut MEDLINE: Exercise Therapy, Telerehabilitation. Die Arbeit wurde bisher 38-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union (FESR), Italy, Italian Ministry of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In recent years, the use of inertial-based systems has been applied to remote rehabilitation, opening new perspectives for outpatient assessment. In this study, we assessed the accuracy and the concurrent validity of the angular measurements provided by an inertial-based device for rehabilitation with respect to the state-of-the-art system for motion tracking. Data were simultaneously collected with the two systems across a set of exercises for trunk and lower limbs, performed by 21 healthy participants. Additionally, the sensitivity of the inertial measurement unit (IMU)-based system to its malpositioning was assessed. Root mean square error (RMSE) was used to explore the differences in the outputs of the two systems in terms of range of motion (ROM), and their agreement was assessed via Pearson’s correlation coefficient (PCC) and Lin’s concordance correlation coefficient (CCC). The results showed that the IMU-based system was able to assess upper-body and lower-limb kinematics with a mean error in general lower than 5° and that its measurements were moderately biased by its mispositioning. Although the system does not seem to be suitable for analysis requiring a high level of detail, the findings of this study support the application of the device in rehabilitation programs in unsupervised settings, providing reliable data to remotely monitor the progress of the rehabilitation pathway and change in patient’s motor function.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Exercise TherapyTelerehabilitationHumansBiomechanical PhenomenaExerciseRange of Motion, Articular

Stichwörter der Autor:innen

exercise therapyfunctional assessmentinertial sensorsmotion analysisrehabilitationtele-rehabilitation

Autor:innen und Institutionen

  • Department of Electronics, Information and Bioengineering, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milano, Italy.
  • Orthopaedic Rehabilitation Unit and Research Laboratory in Biomechanics, Rehabilitation and Ergonomics, San Giuseppe Hospital, IRCCS Istituto Auxologico Italiano, Strada Luigi Cadorna 90, 28824 Piancavallo, Italy.
  • Department of Surgical Sciences, Physical Medicine and Rehabilitation, University of Turin, 10126 Turin, Italy.
  • Clinica Hildebrand, Centro di Riabilitazione Brissago, Via Crodolo 18, 6614 Brissago, Switzerland.
  • Euleria Health Società Benefit Rovereto, 38068 Trento, Italy.

ORCID-Profile: Serena Cerfoglio, Paolo Capodaglio, Paolo Rossi, Ilaria Conforti, Manuela Galli, Veronica Cimolin

Förderung

  • European Union (FESR), Italy
  • Italian Ministry of Health

Zitieren

APA

Cerfoglio, S., Capodaglio, P., Rossi, P., Conforti, I., D'Angeli, V., Milani, E., Galli, M., & Cimolin, V. (2023). Evaluation of Upper Body and Lower Limbs Kinematics through an IMU-Based Medical System: A Comparative Study with the Optoelectronic System. Sensors, 23(13), 6156. https://doi.org/10.3390/s23136156

Vancouver

Cerfoglio S, Capodaglio P, Rossi P, Conforti I, D'Angeli V, Milani E, et al. Evaluation of Upper Body and Lower Limbs Kinematics through an IMU-Based Medical System: A Comparative Study with the Optoelectronic System. Sensors. 2023;23(13):6156. doi:10.3390/s23136156

BibTeX
@article{Cerfoglio2023,
  author = {Cerfoglio, Serena and Capodaglio, Paolo and Rossi, Paolo and Conforti, Ilaria and D'Angeli, Valentina and Milani, Elia and Galli, Manuela and Cimolin, Veronica},
  title = {Evaluation of Upper Body and Lower Limbs Kinematics through an IMU-Based Medical System: A Comparative Study with the Optoelectronic System},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {13},
  pages = {6156},
  doi = {10.3390/s23136156}
}
Kostenlos starten →